A Study on the Physical Properties of Addition Silicone Impression Materials

부가중합형실리콘인상재의 물리적 특성에 대한 연구

  • Kim, Soo-Hwa (Department of Dental Hygiene, Hanyang woman's College)
  • 김수화 (한양여자대학 치위생과)
  • Received : 2006.10.23
  • Accepted : 2006.12.11
  • Published : 2006.12.30

Abstract

This study evaluated the properties of polyvinylsiloxane impression materials: Strain-in-compression, Elastic recovery, Consistency, and Wettability. Four polyvinylsiloxane impression materials(Examixfine, GC, Japan; Aquasilultra, Dentsply, USA; Extrude, Kerr, USA; Perfect-F, Handae, Korea) were used in this study. The measured data were analyzed with SPSSWIN 12.0 program. The results were as follows : 1. The strain-in-compression values showed that Extrude had the lowest value, 4.86% and Examixfine had the highest value, 6.78%. Significant difference between Examixfine and Perfect-F was not found but the other groups had significant differences(p = 0.000). 2. In the elastic recovery test, Extrude had the highest, 99.83% and Perfect-F had the lowest, 96.54%. There is no significant difference between Examixfine and Aquasilultra. The other impression material groups showed significant differences(p = 0.000). 3. The measurement of consistency results that the diameter of Examixfine was the largest, 45.12mm and one of Perfect-F was the smallest, 40.28 mm. No differences were shown between Examixfine and Aquasilultra and between Extrude and Perfect-F(p = 0.001). 4. Aquasilultra had a better hydrophilicity($47.85^{\circ}$) than other materials. Perfect-F had a larger contact angle($94.89^{\circ}$) and was hydrophobic. There were significant differences between groups(p = 0.000). 5. As for correlation among the test types, contact angle correlated with strain-in-compression, elstic recovery and consistency. The strain-in-compression had correlation to contact angle and elastic recovery. The elastic recovery correlated with contact angle and strain-in-compression. And the consistency had correlation to contact angle.

본 연구는 부가중합형실리콘인상재의 정밀 인상채득에 영향을 미치는 요인, 압축변형률, 탄성회복률, 점주도와 젖음성을 평가하기 위하여 현재 시판되고 있는 4종류의 인상재 Examixfine(GC), Aquasilultra(Dentsply), Extrude(Kerr), Perfect-F(Handae)를 이용하여 실험하였다. 실험 데이터는 SPSSWIN 12.0을 사용하여 분석하였고, 결과는 다음과 같다. 1. 압축변형률은 Extrude가 4.86%로 가장 낮게 나타났고, Examixfine이 6.78%로 가장 크게 나타났다. Examixfine과 Perfect-F 사이에는 유의한 값의 차이가 없었으나, 다른 인상재 그룹에서는 유의성이 있었다. (p < 0.001). 2. Extrude의 탄성회복률이 99.83%로 가장 크게 나타났고, Perfect-F가 96.54%로 가장 낮게 나타났다. Examixfine과 Aquasilultra 간에는 유의성이 없었으나, 다른 인상재그룹간에는 유의한 차이가 있었다(p < 0.001). 3. 점주도의 측정결과에서 Examixfine이 점주도 값이 45.12mm로 가장 높았고, Perfect-F의 경우 40.28 mm로 가장 낮은 점주도 값을 나타내었다. Examixfine과 Aquasilultra, Extrude와 Perfect-F 간에는 유의한 차이가 나타나지 않았다(p > 0.001). 4. 젖음성에서는 Aquasilultra가 $47.85^{\circ}$로 가장 접촉각이 작게 나타나 친수성이 가장 우수하게 나타났고, Perfect-F는 접촉각이 $94.89^{\circ}$로 접촉각이 크게 나타나 소수성으로 나타났다. 인상재 종류별로도 각각 유의성이 존재하였다 (p < 0.001). 5. 접촉각은 압축변형률(p < 0.01), 탄성회복률(p < 0.01), 점주도(p < 0.05)와 상관관계가 있었고, 탄성회복률과 점주도 와는 음의 상관관계를 보였다. 압축변형률은 접촉각, 탄성회복률과 상관관계가 있었고(p < 0.01), 탄성회복률과 음의 상관관계를 나타냈다. 탄성회복률은 접촉각, 압축변형률과 음의 상관관계를 가지고 있었고(p < 0.01), 점주도도 접촉각과 음의 상관관계를 보였다(p < 0.05).

Keywords