An Experimental Study for the Prevention of CT Contrast Media Extravasation with Extravasation Detection Accessory System in Femoral Vein of Rabbit

가토 정맥에서 CT 조영제의 혈관외유출 예방을 위한 EDA 시스템의 실험적 연구

  • Kweon, Dae-Cheol (Departments of Diagnostic Radiology, Seoul National University Hospital) ;
  • Jeong, Seok-Hee (Departments of Biomedical Engineering, Seoul National University Hospital) ;
  • Yang, Sung-Hwan (Department of Prosthetics and Orthotics, Korean National College of Rehabilitation and Welfare) ;
  • Cho, Mun-Son (Department of Prosthetics and Orthotics, Korean National College of Rehabilitation and Welfare) ;
  • Jang, Keun-Jo (Department of Diagnostic Radiology, Presbyterian Medical Center) ;
  • Kim, Sun-Geun (Department of Diagnostic Radiology, Woosuk University Oriental Medicine Hospital) ;
  • Yoo, Beong-Gyu (Department of Radiotechnology, Wonkwang Health Science College) ;
  • Lee, Jong-Seok (Department of Radiotechnology, Wonkwang Health Science College)
  • 권대철 (서울대학교병원 진단방사선과) ;
  • 정석희 (서울대학교병원 의공학과) ;
  • 양성환 (한국재활복지대학 의료보장구과) ;
  • 조문선 (한국재활복지대학 의료보장구과) ;
  • 장근조 (예수병원 영상의학과) ;
  • 김순근 (우석대학교 부속 한방병원 방사선과) ;
  • 유병규 (원광보건대학 방사선과) ;
  • 이종석 (원광보건대학 방사선과)
  • Published : 2006.12.31

Abstract

To assess the ability of an extravasation defection accessory (EDA) to detect the clinically important extravascular Injection of iodinated contrast material that was delivered with an automated mechanical power injector. The purpose of this study was to assess the ability of this device during clinically important episodes of extravasation. The EDA system was composed of a strain gage, an amplifier and a computer-based system. In the rabbit experimental cases, there were seven true-positive cases (range of the extravasation volumes: $14{\sim}23 ml$). The algorithm results showed seven true-positive cases (range of the extravasation volumes: $7{\sim}16ml$), nineteen true-negative cases, two false-positive cases and no false-negative cases. The EDA system had a sensitivity of 100% and a specificity of 90% for the detection of clinically important extravasation. The EDA system had good sensitivity for the detection of clinically important extravasation and the EDA system has the clinical potential for the early detection of extravasation of the contrast medium that is administered with power injectors.

CT 검사에서 조영제 혈관외유출을 예방하기 위해 EDA 시스템을 개발하여 가토를 대상으로 실험적으로 혈관외유출의 검출 효율을 평가하였다. 7마리의 가토 IV 정맥 부위에 EDA 시스템의 스트레인게이지를 부착하여 조영제를 주입하면서 혈관외유출 검출 효율을 평가하였다. 조영제의 혈관외유출은 7건으로 혈관외유출 용량 $14{\sim}23ml$ (평균 18.86 ml), 혈관외유출 발생시간 평균 11.29초, 평균압력 109.71 psi, EDA 시스템에서 스트레인게이지의 분해능 평균 512.18, 진양성 7건, 진음성 19건, 위양성 2건이었고, 민감도 100%, 특이도 90%로 나타났다. 개발된 EDA 시스템을 가토에 실험적으로 적용하여 소량의 혈관외유출 조영제를 검출하는 효율이 있어 혈관외유출을 예방하는 효과가 있다.