The Development of Air-kerma Strength Calculation Algorithm in Terms of the Absorbed Dose to Water for HDR Ir-192 Source
기준점에서의 물 흡수선량을 이용한 Ir-192 선원의 공기커마 세기 계산을 위한 알고리즘 개발
- Huh, Hyun-Do (Department of Radiation Oncology, College of Medicine, Inha University) ;
- Kim, Woo-Chul (Department of Radiation Oncology, College of Medicine, Inha University) ;
- Loh, John-Jk (Department of Radiation Oncology, College of Medicine, Inha University) ;
- Lee, Suk (Department of Radiation Oncology, College on Medicine) ;
- Lee, Sang-Hoon (Department of Medical Physics Kyonggi University) ;
- Cho, Sam-Ju (Department of Medical Physics Kyonggi University) ;
- Shin, Dong-Oh (Department of Radiation Oncology, Medicine, College of Medicine, KyungHee University) ;
- Choi, Jin-Ho (Department of Radiation Oncology, Gacheon Medical School, Gil Medical Center) ;
- Kwon, Soo-Il (Department of Medical Physics Kyonggi University) ;
- Kim, Seong-Hoon (Department of Radiation Oncology, College of Medicine, Hanyang University)
- 허현도 (인하대학교 의과대학 방사선종양학교실) ;
- 김우철 (인하대학교 의과대학 방사선종양학교실) ;
- 노준규 (인하대학교 의과대학 방사선종양학교실) ;
- 이석 (고려대학교 의과대학 방사선종양학교실) ;
- 이상훈 (경기대학교대학원 의학물리학과) ;
- 조삼주 (경기대학교대학원 의학물리학과) ;
- 신동오 (경희대학교 의과대학 방사선종양학교실) ;
- 최진호 (가천의과학대학교 방사선종양학교실) ;
- 권수일 (경기대학교대학원 의학물리학과) ;
- 김성훈 (한양대학교 의과대학 방사선종양학교실)
- Published : 2006.12.31
Abstract
The aim of this study was to develop the calculation algorithm of source strength of Ir-192 source In terms of the absorbed dose to water instead of an apparent activity (Ci). For this work the Multi Purpose Brachytherapy Phantom(MPBP) was developed, which was designed to locate the source and the chamber precisely at a specific position Inside the water phantom. The reference point of measurement was set at the 5 cm distance along the transverse axis of the source. For a brachytherapy source calibration, the absorbed dose to water calibration factor (
본 연구의 목적은 Ir-192 선원의 검교정을 겉보기 방사능(apparent activity)을 이용하지 않고 물속 기준점에서 흡수선량을 측정함으로써 공기커마 세기(air-kerma strength, Sk)를 계산하고자 알고리즘을 개발하였다. 연구를 위하여 근접치료용 다목적 팬톰(multi purpose brachytherapy Phantom, MPBP)을 제작하였다. 물 흡수선량 측정은 고선량률 근접치료기(micro-Selectron, Nucletron, Netherlands)에 장착된 Ir-192 선원(Mallinckrodt Medical B.V., Netherlands)을 대상으로 측정하였다. 물 흡수선량은 몬테칼로 계산방법으로 계산된 이온전리함(TM30013, PTW, Germany)의 물 흡수선량 교정인자(