플라즈마 광중합기가 복합레진 중합에 미치는 영향

Effects of plasma arc curing lights on the surface hardness of the composite resins

  • 이수원 (전남대학교 치과대학 소아치과학교실 및 치의학 연구소) ;
  • 양규호 (전남대학교 치과대학 소아치과학교실 및 치의학 연구소) ;
  • 김선미 (전남대학교 치과대학 소아치과학교실 및 치의학 연구소) ;
  • 최남기 (전남대학교 치과대학 소아치과학교실 및 치의학 연구소)
  • Lee, Soo-Won (Department of Pediatric Dentistry, College of Dentistry, Chonnam National University and Dental Research Institute) ;
  • Yang, Kyu-Ho (Department of Pediatric Dentistry, College of Dentistry, Chonnam National University and Dental Research Institute) ;
  • Kim, Seon-Mi (Department of Pediatric Dentistry, College of Dentistry, Chonnam National University and Dental Research Institute) ;
  • Choi, Nam-Ki (Department of Pediatric Dentistry, College of Dentistry, Chonnam National University and Dental Research Institute)
  • 발행 : 2006.11.30

초록

기존에 사용하고 있는 할로겐 광중합기는 여러 가지 장점에도 불구하고 중합시간이 오래 걸린다는 문제 때문에 시술시간이 길어지는 단점이 있는데, 최근에 개발된 플라즈마 광중합기는 매우 짧은 시간에 중합시킬 수 있다고 제조회사는 주장하고 있다. 본 연구에서는 임상에서 흔히 상용하고 있는 복합 레진을 광중합 할 때, 할로겐 광중합으로 얻을 수 있는 플라즈마 광중합기의 적절한 중합시간을 알아보고자 한다. 2mm 두께의 레진 샘플을 만들어 광중합 하고 24 시간 후 상, 하면의 미세경도를 측정하였다. point $4^{(R)}$에서는 플라즈마 광중합기로 6초간 중합했을 때 할로겐과 유사한 경도를 얻었지만, $Z250^{(R)}$에서는 $Flipo^{(R)}$만 9초간 중합시 할로겐과 유사한 광중합을 나타냈다. 이번 연구에서 사용한 플라즈마 광중합기는 적어도 6초 혹은 9초정도 광중합 했을 때 할로겐과 유사한 중합을 얻을 수 있다는 것을 알 수 있었다.

In recent years, xenon plasma arc lamp was introduced for high-intensity curing of composite filling materials in direct resin restorations. In this study, two types of restorative materials, namely composites point $4^{(R)}$ and $Z250^{(R)}$ were selected and curing was conducted using a conventional halogen light and two plama curing lights. Two different resin composites were cured using the different units($Flipo^{(R)}$, Ultra-lite 180A, and $TriLight^{(R)}$) and tested for microhardness. The purpose of this study was to test the hypothesis that exposure to a plasma curing lamp for 3, 6. 9 seconds is equivalent to 20 or 40 seconds of irradiation using a conventional halogen curing unit. 1. $Flipo^{(R)}$ and Ultra-lite 180A were able to polymerize point $4^{(R)}$ at 6 seconds to a degree equal to that of the $TriLight^{(R)}$(control) at 40 seconds. 2. $Flipo^{(R)}$ was able to polymerize $Z250^{(R)}$ at 9 seconds to a degree equal to that of the $TriLight^{(R)}$(control) on the bottom surface at 20 seconds. whereas Ultra-lite 180A could not do. 3. Two plasma curing units were able to cure the test-composites with bottom/top ratios approximately 61% to 96% at 3 to 9 seconds. There were some differences between the two composite brands, with $Z250^{(R)}$ displaying less difference between top and bottom hardness values. For point $4^{(R)}$ and $Z250^{(R)}$, at least 6 or 9 seconds were necessary to produce microhardness equivalent to that of the $TriLight^{(R)}$ curing at 20 or 40 seconds.