Abstract
It is judged that there could be certainty in every process of analyzing environmental samples using Liquid Scintillation Counter. Therefore, this study focuses on quantitative evaluation on uncertainty in an effort to analyze comparatively accurately environmental samples. For this, after parameters which can have an effect on uncertainty was derived, the evaluation on each parameter was tamed out. The results of analysis of each parameter showed that the effect according to the weight difference of Teflon vial did not appear, and that standard deviations of SQP(E) averages reached saturation point at $75{\sim}90$ sec at the result of making increases step by step the irradiation time of External standard, and that values measured by repeat method produces good results compared with replicate. Also, conclusion was derived that analysis on sample after it is left in cold and dark room at least above 1,000 minutes have to be carried out, and the result of carrying out verification on results measured as well as equipment itself using radioactivity-error-analysis and chi-square test, resonable result was derived.
액체섬광계수기를 이용하여 환경시료를 분석하는 모든 과정에서 계측 조건 및 방법론 차이에 의한 불확실성이 존재할 수 있을 것으로 판단하여 본 연구에서는 이에 대한 평가를 정량적으로 수행함으로써 비교적 정확하게 환경시료를 분석하고자 하였다. 이를 위해 결과에 영향을 미칠 수 있는 변수를 도출한 후 각 변수에 대한 영향을 평가한 결과, 분석시료를 제조하는 과정에서의 인적 및 물리적인 불확실성은 무시할 정도로 미미하였고, 측정용기의 무게 차이에 따른 영향은 나타나지 않았다. 외부선원의 조사시간을 단계적으로 증가시킨 결과 시간이 증가함에 따라 데이터의 분산 정도는 감소하면서 $75{\sim}90$ sec에서 포화상태에 도달하였고, Repeat 방법이 Replicate 방법에 비해 데이터의 신뢰성이 높게 나타났다. 또한 방치시간에 따른 영향을 평가한 결과 분석 냉 암소에서 약 1,000 min 이상 방치시킨 후 시료에 대한 분석을 수행해야만 잔상 및 이상 유동에 의한 영향은 거의 없는 것을 알 수 있었고, 결과에 대한 검증을 수행하기 위해 방사능 오차분석과 함께 Chi-square test를 수행한 결과 신뢰성 있는 결과를 보여주었으며 이러한 분석 및 검증 결과에 근거하여 계측 결과에 대한 오차 및 불확실성을 감소시킬 수 있었다.