The Journal of Natural Sciences (자연과학논문집)
- Volume 16 Issue 1
- /
- Pages.31-37
- /
- 2005
- /
- 1225-6196(pISSN)
Determination of the Decenter Position of a Test Lens in a Six-Sided Prism Lens Lensmeter
6면 프리즘 렌즈가 장착된 Lensmeter에서 시험 렌즈의 편심 위치 계산
- Lin, Maria (Dept. of Computational Electronical Physics, Pai-Chai Univ.) ;
- Park, Jong-Dae (Dept. of Computational Electronical Physics, Pai-Chai Univ.) ;
- Jo, Chang-Ho (Dept. of Computational Electronical Physics, Pai-Chai Univ.) ;
- Kim, Hyeon-Gyu (Topins)
- Published : 2005.12.30
Abstract
We have developed an algorithm to determine the position from the image points in a lensmeter with 6-sided prism lens. The positions of the image points which are formed by six prisms depends on the decenter position of a test lens and can be calculated by skew ray tracing. The optical characteristics of the lens meter was analyzed using the Code V program and the positions of the image points was expressed as a function of the decenter position of the test lens. By minimizing the expectation error, we can determine the decenter position position of a test lens from the image positions.
6면 프리즘 렌즈가 장착된 렌즈메터에서 상점의 위치로부터 시험렌즈의 편심의 위치를 알아 낼 수 있는 알고리즘을 개발하였다. 6개의 프리즘에 의해 형성되는 6개의 상점은 시험렌즈의 편심의 극좌표에 의해 변하게 되며, 비틀린 광선 추적으로 계산할 수 있다. 광학 설계 프로그램인 Code V를 사용하여 렌즈메터 광학계를 분석하고, 시험렌즈의 위치에 따른 상점의 위치 변화를 함수로 나타내서 역으로 상점의 위치로부터 시험렌즈의 편심의 위치를 알아낼 수 있는 방법을 개발하였다.