DOI QR코드

DOI QR Code

Phase transition features of binary Co-C eutectic temperature fixed-point

이원계 Co-C 공정계 온도 고정점의 특성

  • Kim, Yong-Gyoo (Temperature-Humidity Group, Div. of Physical Metrology, Korea Research Institute of Standards and Science (KRISS)) ;
  • Yang, In-Seok (Temperature-Humidity Group, Div. of Physical Metrology, Korea Research Institute of Standards and Science (KRISS)) ;
  • Gam, Kee-Sool (Temperature-Humidity Group, Div. of Physical Metrology, Korea Research Institute of Standards and Science (KRISS))
  • 김용규 (한국표준과학연구원 물리표준부 온습도그룹) ;
  • 양인석 (한국표준과학연구원 물리표준부 온습도그룹) ;
  • 감기술 (한국표준과학연구원 물리표준부 온습도그룹)
  • Published : 2005.11.30

Abstract

A Co-C eutectic cell for thermocouple calibration was manufactured and tested to investigate its phase transition characteristics using Type B thermocouples. It was observed that the freezing plateaus were flatter than those of melting, but the melting points were closer to the true transition temperature than the freezing points. The expanded uncertainty of melting temperature was calculated not to exceed $0.2^{\circ}C$ (k = 2). Based on the observed results, the melting process is recommended for the calibration of thermocouples.

Keywords

References

  1. H. Preston-Thomas, 'The international temperature scale of 1990 (ITS-90)', Metrologia, vol. 27, pp. 3-10, 1990 https://doi.org/10.1088/0026-1394/27/1/002
  2. R. E. Bedford, G Bonnier, H. Mass, and F. Pavese, 'Recommended values of temperature on the ITS90 for a selected set of secondary reference points', Metrologia, vol. 33, pp. 133-154, 1996 https://doi.org/10.1088/0026-1394/33/2/3
  3. Y.-G. Kim, K. S. Gam, and K. H. Kang, 'Realization of the palladium freezing point for thermocouple calibrations', Metrologia, vol. 36, pp. 465, 1999 https://doi.org/10.1088/0026-1394/36/5/8
  4. Y. Yamada, H. Sakate, F. Sakuma, and A. Ono, 'Radiometric observation of melting and freezing plateaus for a series of metal-carbon eutectic points in the range $1330^{\circ}C$ to $1950^{\circ}C$', Metrologia, vol. 36, pp. 207-209, 1999 https://doi.org/10.1088/0026-1394/36/3/6
  5. Y. Yamada, F. Sakuma, and A. Ono, 'Thermocouple observations of melting and freezing plateaus for metal-carbon eutectics between the copper and palladium points', Metrologia, vol. 37, pp. 71-73, 2000 https://doi.org/10.1088/0026-1394/37/1/10
  6. Y. Yamada, H. Sakate, F. Sakuma, and A. Ono, 'High-temperature fixed points in the range $1150^{\circ}C$ to $2500^{\circ}C$ using metal-carbon eutectics', Metrologia, vol. 38, pp. 213-219, 2001 https://doi.org/10.1088/0026-1394/38/3/3
  7. M. Sadli, J. Fischer, Y. Yamada, V. Sapritsky, D. Lowe, and G. Machin, 'Review of metal-carbon eutectic temperatures: proposal for new ITS-90 secondary points', Proc. TEMPMEKO 2004 (Cavtat-Dubrovnik), 2004
  8. R. Morice, M. Megharfi, J.-O. Favreau, E. Morel, I. Didialaoui, and J-R. Filtz, 'Realization of metalcarbon fixed points for calibration of contact thermometers at high temperatures', Proc. TEMPMEKO 2004 (Cavtat-Dubrovnik), 2004
  9. CCT/05-21, 'Working Group 2 Report to the CCT', BIPM, 2005
  10. Y.-G. Kim, K. S. Gam, and K. H. Kang, 'A nickel freezing-point cell for thermocouple calibration', Metrologia, vol. 38, pp. 319-323, 2001 https://doi.org/10.1088/0026-1394/38/4/5
  11. 측정불확도 표현지침, 한국표준과학연구원, KRISS99-070-SP, 1999