Korean Journal of Clinical Laboratory Science (대한임상검사과학회지)
- Volume 36 Issue 2
- /
- Pages.137-143
- /
- 2004
- /
- 1738-3544(pISSN)
- /
- 2288-1662(eISSN)
Comparative Evaluation of Erythrocyte Sedimentation Rate by the Test 1 Analyzer and Westergren method in Patients with Anemia
빈혈환자에서 자동 ESR 기기인 Test 1과 Westergren법의 비교
- Kim, Il Taek (Department of Clinical Pathology, Daejeon Health College) ;
- Kim, Jong Seok (Department of Laboratory medicine, Chungbuk Nat'l Univ. Hospital) ;
- Park, Yeon Bo (Department of Laboratory medicine, Chungnam Nat'l Univ. Hospital) ;
- Lim, Jae Sil (Department of Laboratory medicine, Chungnam Nat'l Univ. Hospital)
- Published : 2004.12.31
Abstract
Erythrocyte sedimentation rate(ESR) measurements were performed in a total of 191 blood samples from patients below a hematocrit of 35% by both the Test-1(SIRE Analytical Systems, Udine, Italy) and the Westergren method, endorsed as the reference method for ESR by the International Council for Standardization in Hematology(ICSH). The corrected Westergren ESR values were also obtained applying the formula of Fabry (corrected ESR = ESR measured x 15/55-Hct). Linear regression analysis showed a close correlation (r=0.85) between the two methods both before and after ESR correction in total samples. There was no significant correlation difference between two methods in both two groups with hemoglobin