Abstract
A new vacuum assisted auto-lancing technique is proposed to minimize pain. Specially designed lancing device was introduced, which applied -100mmHg right after skin puncture on the forearm. Sampled blood volumes were measured in 58 normal females. Mean volume of 464 samples was approximately 2.6$\muL$ and the frequency of more than 0.5$\muL$ was 86%. Thus the success rate of blood sugar test should also be the same when using modern glucose meters capable of testing with only 0.3~0.5$\muL$ of capillary blood. When pain scores were quantitatively evaluated by the visual pain measure, only 23% pain of the traditional finger sampling was experienced, demonstrating that capillary blood sampling was performed on the forearm with almost no pain. The present technique reduced pain to a great degree, though resampling might be unavoidable due to 14% of test failure rate estimated for modern glucose meters. However, minimized pain makes the present technique of great convenience for diabetic patients who need blood sampling a few times a day.
당뇨병 환자의 혈당검사시 통증을 최소화하는 새로운 진공 자동 채혈기법을 개발하였다. 특별히 고안, 제작된 채혈기구를 사용하여 전완부위에서 채혈침으로 피부를 관통한 직후 자동적으로 -100mmHg의 진공압력이 가해짐으로써 모세혈액이 흡입되는 자동화된 진공 채혈기법을 구현하였다. 진공 자동 채혈기법의 유용성을 검증하기 위하여 전완부위에서의 채혈용적을 측정하였다. 58명의 정상인을 대상으로 464회의 채혈을 시행한 결과 평균 채혈용적은 약 2.6$\muL$이었고 0.5$\muL$이상의 채혈빈도가 약 86%이었다. 최근 개발되고 있는 혈당계들은 0.3~0.5$\muL$의 모세혈액으로 혈당검사가 가능하므로 검사성공률 역시 약 86%로 추정되었다. 채혈통증 감소효과를 평가하기 위하여 손가락 및 전완부위 채혈 후 시각상사 통증척도를 사용하여 통증지수를 측정하였다. 전완부위 채혈시 통증은 손가락 채혈의 약 23%에 불과하였으며 거의 무통증 채혈이 가능하였다. 따라서 본 연구의 전완부위 진공 자동채혈기법은 전통적인 손가락 채혈에 비해 채혈통증이 25% 이하로 현저하게 감소하지만 검사실패율이 14%로써 경우에 따라 재채혈이 필요할 수 있다. 그러나 매일 수 차례 채혈해야 하는 만성 당뇨환자의 경우 검사기피를 방지할 수 있는 유용한 방법으로 생각된다.