Tc-99m ECD 뇌혈류 SPECT 시행시 부적절한 영상획득 모수에 의한 인공산물

An Artifact Caused by Using the Inadequate Acquisition Parameter in Tc-99m ECD Brain Perfusion SPECT

  • 서한경 (전북대학교 의과대학 핵의학교실) ;
  • 김민우 (전북대학교 의과대학 핵의학교실) ;
  • 임석태 (전북대학교 의과대학 핵의학교실) ;
  • 손명희 (전북대학교 의과대학 핵의학교실)
  • Seo, Han-Kyung (Departments of Nuclear Medicine, Chonbuk National University Medical School) ;
  • Kim, Min-Woo (Departments of Nuclear Medicine, Chonbuk National University Medical School) ;
  • Lim, Seok-Tae (Departments of Nuclear Medicine, Chonbuk National University Medical School) ;
  • Sohn, Myung-Hee (Departments of Nuclear Medicine, Chonbuk National University Medical School)
  • 발행 : 2004.12.31

초록

Although artifacts in tomographic imaging can arise from a number of sources, this case presents an artifact caused by using inadequate acquisition parameters which do not match the collimator with which a camera equipped in Tc-99m ECD brain SPECT. Whenever the collimator in tomographic studies is changed, we should take care to use suitable parameters of acquisition protocol according to the collimator with which a camera is fitted. Other ways to prevent an error of this type is to use a point source.

키워드

참고문헌

  1. Wackers FJ. Artifacts in planar and SPECT myocardial perfusion Imaging. Am J Card Imaging 1992;6:42-57
  2. Shih WJ, Pulmano C, Han JK. Artifact of cold lesions on skull SPECT of a patient with Paget's disease secondary to pixel overflow. Clin Nucl Med 1999;24:372-3 https://doi.org/10.1097/00003072-199905000-00025
  3. Lange G, Van Niekerk A, Meyer BJ. Detection of an artifact on lumbar SPECT. Clin Nucl Med 2001;26:446-8 https://doi.org/10.1097/00003072-200105000-00014
  4. Depuey EG III. How to detect and avoid myocardial perfusion SPECT artifacts. J Nucl Med 1994;35:699-702
  5. Depuey EG, Garcia EV. Optimal specificity of thallium-201 SPECT through recognition of imaging artifacts. J Nucl Med 1989;30:441-9
  6. Zhang JJ, Park CH, Kim SM. Brain SPECT artifact in multidetector SPECT system. Clin Nucl Med 1994;19:789-91 https://doi.org/10.1097/00003072-199409000-00009
  7. Hackett MT, Shih WJ, Kiefer VH. Brain SPECT artifacts in patients having metallic cranioplasty. J Nucl Med Technol 2001;29:189-92
  8. Boren EL, Cowan CL, Anderson SG. First impressions: brain SPECT with cortical photopenic defect. J Nucl Med 1996;37:1165
  9. Zhang JJ, Park CH, Kim SM, Intenzo C, Reyes P. The attenuation effect in brain SPECT. Clin Nucl Med 1993;18:583-6 https://doi.org/10.1097/00003072-199307000-00009