Nitrogen Wash-Out Technique to Measure Functional Residual Capacity Based on Expired o2/Co2 Analysis

o2/Co2 분석기를 사용하여 폐의 기능적 잔기용량을 계측하는 질소세척법 개발

  • Kim, Goon-Jin (Cheongju Polytechnic College) ;
  • Kim, Kyung-Ah (CKInternational Co., Ltd.) ;
  • Lee, Jae-Hun (CKInternational Co., Ltd.) ;
  • Lee, Tae-Soo (Biomedical Engineering Department, School of Medicine, Chungbuk National University) ;
  • Cha, Eun-Jong (Biomedical Engineering Department, School of Medicine, Chungbuk National University)
  • Published : 2004.09.30

Abstract

Functional residual capacity (FRC) is an important diagnostic parameter measured using $N_{2}$ analyzer. Since $N_{2}$ analyzer is expensive as well as cumbersome for use of noisy vacuum pump, the FRC measurement becomes possible only in large well-equipped hospitals. The present study introduced a new $TN_{2}$ wash-out technique to measure FRC by $O_{2}/CO_{2}$ analysis, which is relatively cheaper and much simpler to apply. Slower $O_{2}$ response was compensated for high frequency to be coincided with $CO_{2}$ response, thereby enabled indirect, but accurate $N_{2}$ concentration measurement. FRC was estimated by continuous integration of expired $N_{2}$ volume obtained with air flow signal. Experiment with 3 L syringe, a standard calibration device recommended by the American Thoracic Society, demonstrated less than 1% error at 0, 1, and 2 L. Correlation coefficient was almost ideal, guaranteeing linear estimation of FRC. The present technique is inexpensive and simple to apply, thus should he of great convenience.

Keywords

References

  1. W. F. Ganong, 'Pulmonary function', In: Review of medical physiology, Section VII. pp. 507-509, LANGE Medical Publications, Los Altos, U. S. A., 1981
  2. F. S. Grodins and S. M. Yamashiro, 'Respiratory function of the lung and its control', pp. 8-16, Macmillan Publishing Co., Inc., New York, 1978
  3. 차은종, '호흡기 시스템의 측정', In: 의용생체공학, 제 9장, pp. 509-598, 의공학교육연구 회 역편, 여문 각, 1993
  4. T. D. East and E. A. East, 'Nitrogen analyzers', In: Encyclopedia of medical devices and instrumentation, vol. 3, pp. 2053-2058, Ed. by J. G. Webster, John Wiley & Sons, 1988
  5. American Thoracic Society, 'Standardization of spiromety 1994 update', Am. J. Respir. & Grit. Gare Med., vol. 152, pp. 1107-1136, 1995 https://doi.org/10.1164/ajrccm.152.3.7663792
  6. 차은종, 김경아, 김현식, '호흡 방해 효과를 최소화한 단방향 기류 밸브관', 대한민국 특허 출원 제 10-2002-000839호, 2002