DOI QR코드

DOI QR Code

열전대(thermocouple)를 이용한 Buchanan Plugger 표면의 온도상승 분석

ANALYSIS OF TEMPERATURE RISE ON THE SURFACE OF BUCHANAN PLUGGER USING THERMOCOUPLE

  • 조진숙 (전남대학교 치과대학 보존학교실, 치의학연구소) ;
  • 황윤찬 (전남대학교 치과대학 보존학교실, 치의학연구소) ;
  • 김선호 (전남대학교 치과대학 보존학교실, 치의학연구소) ;
  • 황인남 (전남대학교 치과대학 보존학교실, 치의학연구소) ;
  • 최보영 (전남대학교 공과대학 기계시스템공학부) ;
  • 정영진 (전남대학교 공과대학 기계시스템공학부) ;
  • 정우남 (전남대학교 공과대학 기계시스템공학부) ;
  • 오원만 (전남대학교 치과대학 보존학교실, 치의학연구소)
  • Cho, Jin-Suk (Dept. of Conservative Dentistry, College of Dentistry, DSRI, Chonnam Natl.Uni.) ;
  • Hwang, Yun-Chan (Dept. of Conservative Dentistry, College of Dentistry, DSRI, Chonnam Natl.Uni.) ;
  • Kim, Sun-Ho (Dept. of Conservative Dentistry, College of Dentistry, DSRI, Chonnam Natl.Uni.) ;
  • Hwang, In-Nam (Dept. of Conservative Dentistry, College of Dentistry, DSRI, Chonnam Natl.Uni.) ;
  • Choi, Bo-Young (School of Mechanical Systems Engineering, Chonnam Natl. Uni.) ;
  • Jeong, Young-Jin (School of Mechanical Systems Engineering, Chonnam Natl. Uni.) ;
  • Juhng, Woo-Nam (School of Mechanical Systems Engineering, Chonnam Natl. Uni.) ;
  • Oh, Won-Mann (Dept. of Conservative Dentistry, College of Dentistry, DSRI, Chonnam Natl.Uni.)
  • 발행 : 2003.07.01

초록

This study was performed to evaluate the actual temperature rise on the surface of Buchanan plugger using thermocouple. The heat carrier system 'System B Heatsource'(Model 1005, Analytic Technologies, Redmond, WA, USA) and the Buchanan pluggers of F, FM, M and ML sizes are used for this study. The temperature was set to 200^{\circ}C on digital display and the power level on it was set to 10. Five thermocouples were placed in direct contact with the surface of each size of Buchanan's pluggers at 1 mm increments from the tip to the 4 mm length of shank. The heat control spring was touched for 5 seconds. and the temperature rise on the surface of the pluggers were measured at 1 sec intervals for more than 5 seconds with an accuracy of 0.01 using Data Logger. The data were statistically analyzed by one-way ANOVA. The results were as follows. 1. The position at which the temperature peaked was approximately at 1~2 mm far from the tip of Buchanan plugger (p<0.01). 2. The peak temperature was $215.25{\pm}2.28^{\circ}C$ in F plugger. $185.94{\pm}2.19^{\circ}C$ in FM plugger, $169.51{\pm}9.12^{\circ}C$ in M plugger, and 160.79{\pm}1.27^{\circ}C in ML plugger and the peak temperature was highest in F plugger and followed by. in descending order. FM plugger. M plugger. ML plugger showed the lowest peak temperature (p<0.01). 3. The temperature on the plugger was decreased with the increase of touching time. This results suggest that the actual temperature on the surface of the pluggers does not correlate well with the temperature set on digital display. Heat concentrates around the tip. The larger plugger reveals lower temperature rise relatively.

키워드

참고문헌

  1. Schilder H : Filling root canals in three dimensions. Dent Clin North Am 11:723-744, 1967
  2. Nguyen NT : Obturation of the Root Canal System. Pathways of the Pulp. 6th ed., St. Louis, Mosby Co., 219-271, 1994
  3. Jurcak JJ, Weller RN, Kulild JC and Donley DL : In vitro intracanal temperatures produced during warm lateral condensation of gutta-percha. J Endod 18:1-3, 1992 https://doi.org/10.1016/S0099-2399(06)81133-7
  4. Lee FS, Van Cura JE and BeGole E : A comparison of root surface temperatures using different obturation heat source. J Endod 24:617-620, 1998 https://doi.org/10.1016/S0099-2399(98)80123-4
  5. Hand RE, Hugel EF and Tsaknis PJ : Effect of a warm gutta-percha technique on the lateral periodontium. Oral Surg 42: 395-401, 1976 https://doi.org/10.1016/0030-4220(76)90174-2
  6. Marlin J and Schilder H : Physical properties of guttapercha when subjected to heat and vertical condensation. Oral Surg 36:872-879, 1973 https://doi.org/10.1016/0030-4220(73)90339-3
  7. Barkhordar RA, Goodis HE, Watanabe L and Koundjian J : Evaluation of temperature rise on the outer surface of teeth during root canal obturation technique. Quint lnt 21:585-588, 1990
  8. Eriksson AR, Albrektsson T, Grane B and McQueen D: Thermal injury to bone : a vital-microscopic description of heat effect. lnt J Oral Surg 11:115-121, 1982 https://doi.org/10.1016/S0300-9785(82)80020-3
  9. Eriksson AR and Albrektsson T : The effect of heat on bone regeneration. Oral Surg 42:705-711, 1984 https://doi.org/10.1016/0030-4220(76)90091-8
  10. Eriksson AR and Albrektsson T : Temperature threshold levels for heat-induced bone tissue injury : a vitalmicroscopic study in the rabbit. J Prosthet Dent 50:101-107, 1983 https://doi.org/10.1016/0022-3913(83)90174-9
  11. Romero AD, Green DB and Wucherpfennig AL : Heat transfer to the periodontal ligament during root obturation procedures using an in vitro model. J Endod 26: 85-87, 2000 https://doi.org/10.1097/00004770-200002000-00006
  12. Weller RN, Jurcak JJ, Donley DL and Kulild JC : A new model system for measuring intracanal temperatures. J Endod 17: 491-494, 1991 https://doi.org/10.1016/S0099-2399(06)81796-6
  13. Floren JW, Weller RN, Pashley DH and Kimbrough WF : Changes in root surface temperatures with in vitro use of the System B HeatSource. J Endod 25:593-595, 1999 https://doi.org/10.1016/S0099-2399(99)80314-8
  14. 최성아, 김선호, 황윤찬, 윤창, 오병주, 최보영, 정우남, 정선와, 황인남과 오원만 : 적외선열화상장치를 이용한 Buchanan plugger 표면의 온도상승 분석, 대한치과보존학회지 27:370-381, 2002
  15. Buchanan LS : Continuous wave of condensation technique. Dentistry Today 1:60-64, 1996
  16. Analytic Technology Corp. Instruction guidelines for system B HeatSource model 1005, Glendora, California, USA, 1997