DOI QR코드

DOI QR Code

양액의 영양분 분석을 위한 Strip형 이온선택성전극 센서 의 응용 -NO3 이온 분석 -

Application of a. Strip Ion-Selective Electrode Sensor for Hydroponic Nutrient Solution Analysis - NO3 Analysis -


초록

A simple disposable thick-film potentiometric strip has been developed and evaluated for hydroponics application. The strip consisted of low ion-selective electrodes (ISE) fabricated by screen-printing technology. The electrochemical responses of ion sensors for nitrate, ammonium, potassium, and magnesium were measured with specially designed 16-channel low voltage signal transducers. The analytical characteristics of the sensors were comparable with those of conventional ISE sensors. The thick-film sensors exhibit linear relationships over five concentration decades. The concentration of N $O_3$ - ion in standard solution can be determined by direct potentiometric measurements without any conditioning before measurements. However, measurement of $K^{+}$, N $H_4$$^{+}$, and $Mg^{2+}$ ionic concentrations in nutrient solutions seems not feasible.

키워드

참고문헌

  1. Modern analytical chemistry Harvey,D.
  2. Anal. Chim. Acta v.302 A potassium-ion selective electrode with valinomycin based poly(vinyl chloride) membrane and a poly(vinyl ferrocene) soild contact Hauser,P.C.;D.W.L.Chiang;G.A.Wright. https://doi.org/10.1016/0003-2670(94)00472-X
  3. Computers and Electronics in agriculture v.14 Soil proterty sensing for site-specific crop management Hummel,J.W.;L.D.Gaultney;K.A.Sudduth https://doi.org/10.1016/0168-1699(95)00043-7
  4. Anal. Chim. Acta v.385 Disposable strip potentiometric electrodes with solvent-polymetric ionselective membranses fabricated using screenprinting technology Koncki,R.;S.Glab;J.Dziwulska;I.Pachetti;M.Mascini https://doi.org/10.1016/S0003-2670(98)00726-0
  5. J. Chromatogr. A v.902 Ion chromatograghy aetector based on solid-state ion-selective electrode array Lee,D.K.;H.J.Lee;G.S.Cha;H.Nam;K.J.Paeng https://doi.org/10.1016/S0021-9673(00)00844-X
  6. Anal. Chim. Acta v.385 Chemical sensor array for multicomponent analysis of biological liquids Legin,A.;A.Smimova;A.Rudniskaya;L.Lvova;E.Suglobova;Y.Vlasov https://doi.org/10.1016/S0003-2670(98)00690-4
  7. Electrochimica v.47 Development of multisensor systems based of chalcogenide thin film chemical sensor for the simultaneous multicomponent analysis of metal ions in complex solutions Mourzina, Yu. G.;J.Schubert;W.Zander;A.Legin;Yu. G. Vlasov;H.Luth;M.J.Schoning https://doi.org/10.1016/S0013-4686(01)00563-1
  8. Application to tobacco anaylsis,Talanta v.53 Construction and evaluation of ion selective electrodes for nitrate with a summing operational amplifier Perex-Olmos, R.;A.Rios;J.R.Femendez;R.A.S.Lapa;J.L.F.C.Lima https://doi.org/10.1016/S0039-9140(00)00526-9
  9. Sensors and Acturators B v.26 Intrument with integraged sensor for a rapid determination of inorganic ions Roβ,B.;G.Steiner;M.Kiesshauer;M.Bradter;K.Cammann
  10. Aldrich Inc. Sigma-Aldrich Hompage.
  11. Sensors and Actuators B v.34 Miniaturised ionselective sensor chip for potassium measurement in a biomedical application Uhlig,A;F.Dietrich;E.Lindner;U.Schnakenberg;R.Hintsche
  12. Sensors and Actuators B v.65 Micromachined glucose sensor and $K^+$ ISE based on containment array Zhu,J.Z.;J.F.Xie;D.R.Lu;G.X.Zhang;Z.R.Zhang https://doi.org/10.1016/S0925-4005(99)00432-3

피인용 문헌

  1. Determination of Inorganic Phosphate in Paprika Hydroponic Solution using a Laboratory-made Automated Test Stand with Cobalt-based Electrodes vol.36, pp.5, 2011, https://doi.org/10.5307/JBE.2011.36.5.326