DOI QR코드

DOI QR Code

INFRARED THERMOGRAPHIC ANALYSIS OF TEMPERATURE RISE ON THE SURFACE OF BUCHANAN PLUGGER

적외선열화상장치를 이용한 Buchanan plugger 표면의 온도상승 분석

  • Choi, Sung-A (Dept. of Conservative Dentistry, College of Dentistry, DSRI, Chonnam Natl. Uni.) ;
  • Kim, Sun-Ho (Dept. of Conservative Dentistry, College of Dentistry, DSRI, Chonnam Natl. Uni.) ;
  • Hwang, Yun-Chan (Dept. of Conservative Dentistry, College of Dentistry, DSRI, Chonnam Natl. Uni.) ;
  • Youn, Chang (Dept. of Conservative Dentistry, College of Dentistry, DSRI, Chonnam Natl. Uni.) ;
  • Oh, Byung-Ju (Dept. of Conservative Dentistry, College of Dentistry, DSRI, Chonnam Natl. Uni.) ;
  • Choi, Bo-Young (School of Mechanical Systems Engineering College of Engineering, Chonnam Natl. Uni.) ;
  • Juhng, Woo-Nam (School of Mechanical Systems Engineering College of Engineering, Chonnam Natl. Uni.) ;
  • Jeong, Sun-Wa (Dept. of Dentistry, College of Medicine, Seonam University) ;
  • Hwang, In-Nam (Dept. of Conservative Dentistry, College of Dentistry, DSRI, Chonnam Natl. Uni) ;
  • Oh, Won-Mann (Dept. of Conservative Dentistry, College of Dentistry, DSRI, Chonnam Natl. Uni)
  • 최성아 (전남대학교 치과대학 보존학교실, 치의학연구소) ;
  • 김선호 (전남대학교 치과대학 보존학교실, 치의학연구소) ;
  • 황윤찬 (전남대학교 치과대학 보존학교실, 치의학연구소) ;
  • 윤창 (전남대학교 치과대학 보존학교실, 치의학연구소) ;
  • 오병주 (전남대학교 치과대학 보존학교실, 치의학연구소) ;
  • 최보영 (전남대학교 공과대학 자동차공학부) ;
  • 정우남 (전남대학교 공과대학 자동차공학부) ;
  • 정선와 (서남대학교 의과대학 치과학교실) ;
  • 황인남 (전남대학교 치과대학 보존학교실, 치의학연구소) ;
  • 오원남 (전남대학교 치과대학 보존학교실, 치의학연구소)
  • Published : 2002.07.01

Abstract

This study was performed to evaluate the temperature rise on various position of the Buchanan plugger, the peak temperature of plugger's type and the temperature change by its touching time of heat control spling. The heat carrier system 'System B' (Model 1005, Analytic Technologies, USA) and the Buchanan's plug-gers of F, FM, M and ML sizes are used for this study. The temperature was set to 20$0^{\circ}C$ which Dr. Buchanan's "continuous wave of condensation" technique recommended on digital display and the power level on it was set to 10. In order to apply heat on the Buchanan's pluggers, the heat control spring was touched for 1, 2, 3, 4 and 5 seconds respectively. The temperature rise on the surface of the pluggers were measured at 0.5 mm intervals from tip to 20 mm length of shank using the infrared thermography (Radiation Thermometer-IR Temper, NEC San-ei Instruments, Ltd, Japan) and TH31-702 Data capture software program (NEC San-ei Instruments, Ltd, Japan). Data were analyzed using a one way ANOVA followed by Duncan's multiple range test and linear regression test. The results as follows. 1. The position at which temperature peaked was approximately at 0.5 mm to 1.5 mm far from the tip of Buchanan's pluggers (p<0.001). The temperature was constantly decreased toward the shank from the tip of it (p<0.001). 2. When the pluggerss were heated over 5 seconds, the peak temperature by time of measurement revealed from 253.3$\pm$10.5$^{\circ}C$ to 192.1$\pm$3.3$^{\circ}C$ in a touch for 1 sec, from 218.6$\pm$5.$0^{\circ}C$ to 179.5$\pm$4.2$^{\circ}C$ in a touch for 2 sec, from 197.5$\pm$3.$0^{\circ}C$ to 167.5$\pm$3.7$^{\circ}C$ in a touch for 3 sec, from 183.7$\pm$2.5$^{\circ}C$ to 159.8$\pm$3.6$^{\circ}C$ in a touch for 4 sec and from 164.9$\pm$2.$0^{\circ}C$ to 158.4$\pm$1.8$^{\circ}C$ in a touch for 5 sec. A touch for 1 sec showed the highest peak temperature, followed by, in descending order, 2 sec, 3 sec, 4 sec. A touch for 5 sec showed the lowest peak temperature (p<0.001). 3. A each type of pluggers showed different peak temperatures. The peak temperature was the highest in F type and followed by, in descending order, M type, ML type. FM type revealed the lowest peak temperature (p<0.001). The results of this study indicated that pluggers are designed to concentrate heat at around its tip, its actual temperature does not correlate well with the temperature which Buchanan's "continuous wave of condensation" technique recommend, and finally a quick touch of heat control spring for 1sec reveals the highest temperature rise.

Keywords

References

  1. Schilder H : Filling root canals in three dimensions. Dent Clin North Am 11:723-744, 1967
  2. Nguyen NT : Obturation of the Root Canal System. Pathways of the Pulp. 6th ed., St. Louis. Mosby Co., 1994. pp 219-271
  3. Hand RE, Hugel EF. Tsaknis PJ : Effect of a warm gutta-percha technique on the lateral periodontium. Oral Surg 42:395-401, 1976 https://doi.org/10.1016/0030-4220(76)90174-2
  4. Marlin J, Schilder H : Physicial properties of guttaperch a when subjected to heat and vertical condensation. Oral Surg 36:872-879, 1973 https://doi.org/10.1016/0030-4220(73)90339-3
  5. Barkhordar RA. Goodis HE, Watanabe L, Koundjian J : Evaluation of teeth during root canal obturation technique. Quint Int 21:585-588. 1990
  6. Eriksson A. Albrektsson T, Grane B, McQueen D : Thermal injury to bone : a vital-microscopic description of heat effect. Int J Oral Surg 11: 115-121, 1982 https://doi.org/10.1016/S0300-9785(82)80020-3
  7. Eriksson AR. Albrektsson T : The effect of heat on bone regeneration. Oral Surg 42: 705-711, 1984 https://doi.org/10.1016/0030-4220(76)90091-8
  8. Eriksson AR. Albrektsson T : Temperature threshold levels for heat-induced bone tissue injury : a vital-microscopic study in the rabbit. J Prosthet Dent 50:101-107,1983 https://doi.org/10.1016/0022-3913(83)90174-9
  9. Lee FS, Van Cura JE, BeGole E : A comparison of root surface temperatures using different obturation heat sources, J Endod 24:617-620, 1998 https://doi.org/10.1016/S0099-2399(98)80123-4
  10. Weller RN. Jurcak JJ, Donley DL, Kulild JC : A new model system for measuring intracanal temperatures. J Endod 17:491-494, 1991 https://doi.org/10.1016/S0099-2399(06)81796-6
  11. Floren JW. Weller RN. Pashley DB, Kimbrough WF :Changes in root surface temperatures with in vitro use of the system B HeatSource. J Endod 25:593-595, 1999 https://doi.org/10.1016/S0099-2399(99)80314-8
  12. McCullagh JJ, Setchell DJ. Gulabivala K. Hussey DL, Biagioni P, Lamey PJ, Bailey G : A comparison of thermocouple and infrared thermographic analysis of temperature rise on the root surface during the continuous wave of condensation technique. Int Endod J 33:326-332, 2000 https://doi.org/10.1046/j.1365-2591.2000.00302.x
  13. Biagioni PA, Longmore RB, McGimpsey JG. Lamey P-J: Infrared thermography - its role in dental research with particular reference to craniomandibular disorders. Dentomaxillofacial Radiology 25: 119-124. 1996 https://doi.org/10.1259/dmfr.25.3.9084259
  14. Buchanan LS : Continuous wave technique. Dentistry today 1:60-64. 1996
  15. Analytic Technology Corp. Instruction guidelines for system B HeatSource model 1005. Redmond, WA: Analytic Technology Corp., 1997
  16. Jurcak JJ. Weller RN. Kulild JC. Donley DL : In vitro intracanal temperatures produced during warm lateral condensation of guttapercha. J Endod 18: 1-3. 1992 https://doi.org/10.1016/S0099-2399(06)81133-7

Cited by

  1. Plugger temperature of cordless heat carriers according to the time elapsed vol.43, pp.1, 2018, https://doi.org/10.5395/rde.2018.43.e12