Flow-Dependent Friction Loss in an Implantable Artificial Lung

  • Lee, Sam-Cheol (Department of Advanced Materials Engineering, Hanlyo University)
  • Published : 2002.11.01

Abstract

The goal of this work is to design and build an implantable artificial lung that can be inserted as a whole into a large vein in the body with the least effect on cardiovascular hemodynamics. The experimental results demonstrate that the pressure drop is not entirely related to viscosity effects. The friction factor decreases with an increase in the number of tied-hollow fibers at a constant Reynolds number A uniform flow pattern without stagnation is observed at all numbers of tied hollow fibers tested. The tied hollow fiber module, built in this study with 3 cm of outer diameter of module. 380 m of outer diameter of tied hollow fiber, and 700 number of tied hollow fiber with length of 60 cm, which shows a pressure drop of 13-16 mmHg, satisfies the required pressure drop qualifying 15 mmHg as an intravascular artificial lung.

Keywords

References

  1. Bartlett, R.H., 1990, 'Extracorporeal Life Support for Cardiopulmonary Failure,' Curr. Probl. Surg, Vol. 27, pp. 623-705 https://doi.org/10.1016/0011-3840(90)90015-W
  2. Bird, R.B., Stewart, W.E. and Light, E.N., 1960, Transport phenomena, New York, Wiley & Song, Inc., New York, pp. 180-201
  3. Borst, H.G. and Schfers, H.J., 1993, 'Lung Transplatation,' Cli. Invest. Vol. 71, pp. 98-101
  4. Egan, T.M., Boychuk, J.E., Rosato, K. and Cooper, J.D., 1992, 'Whence the Lungs? A Study to Asses Suitability of Donor Lungs for Transplanatation,' Transplantation, Vol. 53, pp. 420-422 https://doi.org/10.1097/00007890-199202010-00029
  5. Ergun, S., 1954, 'Fluid Flow through Packed Columns,' Chem. Eng. Prog., Vol. 48, pp. 89-94
  6. Federspiel, William, J., Lund, L.W., Bultman, J.A., Wanant, S., Matoney, J., Golob, J.F., Frankowski, B.J., Watach, M., Litwak, P. and Hattler, B.G., 1999, 'Ex-vivo Testing of the Intravenous Membrane Oxygenator(IMO),' ASAIO J., Vol. 45, No. 2, p. 127
  7. Kirklin, J.K., Westely, S., Blackstone, E.M., Kirklin, J.W., Chenoweth, D.E. and Pacifico, A.D., 1983, 'Complement and the Damaging Effects of Cardiopulmonary Bypass,' J. Thorac. Cardiovasc. Surg. Vol. 86, pp. 845-857
  8. Mortnsen, J.D. and Berry, G., 1989, 'Conceptual and Design Features of a Practical, Clinically Effective Intravenous Mechanical Blood Oxygen/carbon Dioxide Exchange Device (IVOX),' The Intern. Artif. Organs, Vol. 12, No. 6, pp. 384-389
  9. Vaslef, S.N., Mockros, L.F., Cook, K.E., Leonard, R.J., Sung, J.C.and Anderson, R.W., 1994, 'Computer-Assisted Design of an Implantable Intrathoracic Artificial Lung,' Artif. Organs, Vol. 18, No. 11, pp. 813-817 https://doi.org/10.1111/j.1525-1594.1994.tb03328.x