Determination of Iodide in Sulfuric Acid Aqueous Solution by the Ion Chromatography with UV Detection

이온 크로마토그래피와 자외선 검출을 이용한 황산수용액 중의 요오드 음이온 정량

  • Park, Yang-Soon (Nuclear Chemistry Research Team, Korea Atomic Energy Research Institute)
  • 박양순 (한국원자력 연구소. 원자력화학연구팀)
  • Received : 2001.05.02
  • Published : 2001.08.25

Abstract

Ion chromatography was applied to determine iodide remained in sulfuric acid aqueous solution after adsorption procedure. Iodide was determined in 0.25 M, 0.5 M and 1 M sulfuric acid solution on time variation. Because sulfuric acid in solution plays as an oxidizer, the concentration of iodide decreased with increasing concentration or sulfuric acid. Thereafter, sulfuric acid was neutralized with sodium hydroxide. Good linearity(r=0.99999) was obtained at the range of 0-20 mg/L 1 in 0.5 M sodium sulfate matrix.

황상용액 매질에서 요오드 음이온을 흡착제에 흡착시킨 후 황산용액에 남아있는 요오드 음이온을 정량하기 위해 이온 크로마토그래피를 적용하였다. 산화제 작용을 하는 황산의 농도를 0.25 M, 0.5 M, 1 M로 변화시키고 요오드 음이온을 각각 첨가한 후 시간변화에 따라 요오드 음이온이 산화되는 정도를 측정하였다. 황산농도가 진할수록 요오드 음이온은 산화되어 휘발하므로 적게 검출되었다. 수산화나트륨 용액을 사용하여 환산매질을 중화하였다. 중화된 0.5 M $Na_2SO_4$ 매질 중 요오드 음이온의 농도범위가 0-20 mg/L인 구간에서 검정곡선의 직선성이 좋게 나타났다(r=0.99999).