Abstract
In this paper, we have compared the direct fan-beam reconstruction method with the rebinning method in terms of computation time and spatial resolution using computer simulation. As a result, the direct fan-beam method is superior to the rebinning method in the spatial resolution though the former needs longer computation time. However, if we adopt the quarter-detector-offset technique to improve the spatial resolution, the rebinning method outperforms the direct fan-beam method. The computation times have been evaluated using the fast algorithms optimized to reduce the number of interpolation calculations at the back-projection, and the spatial resolutions have been compared using the computer generated phantoms.