Flow Visualizations and Hot-Wire Measurements on Air Flow in Two Different Neonate Incubators

  • Kim, Young-Ho (Department of Biomedical Engineering, College of Health Science, Institute of Medical Instrumentation and Rehabilitation Engineering, Yonsei University) ;
  • Yoo, Seoung-Chool (Department of Mechanical Engineering, Halla University) ;
  • Kwon, Chi-Ho (Department of Biomedical Engineering, College of Health Science, Institute of Medical Instrumentation and Rehabilitation Engineering, Yonsei University)
  • Published : 2001.07.01

Abstract

Flow visualizations and hot-wire measurements on the inside flows of two different incubators are presented in this paper. An anatomically-correct neonate model was fabricated using the rapid prototyping machine, based on the 3-D scanned data. The result showed that air flow in the incubator was affected not only by the air circulation system but also by the design of incubator chamber. Large rotating motions were located around the corners of free space. A number of small eddies were found in regions of high shear flow, in areas such as that between the air inlet and the neonate. But, these small eddies were found to be stationary at that locations. Those small eddies might interfere with convective and evaporative heat transfers from the neonate. This study has led to a better understanding of flow mechanism in an incubator chamber and provided the guidance needed for the advancement of improved computational fluid dynamic models.

Keywords

References

  1. Bell, E. F., 1983, 'Infant Incubators and Radiant Warmers,' Early Hum. Dev., Vol, 8, No. 3-4, pp. 351-375 https://doi.org/10.1016/0378-3782(83)90018-X
  2. Hasegawa, T., Horio, H., Okino, H., Taylor, T. W. and Yamaguchi, T., 1993, 'Three-Dimensional Visualization of Air Flow in Infant Incubators Using Computational Fluid Mechanics,' Biomedical Instrumentation & Technology, Vol, 27, No. 4. pp. 311-317
  3. Hey, E. N., 1969, 'The Relation Between Environmental Temperature and Oxygen Consumption in the Newborn Baby,' J. Physiol, Vol, 200, No. 3, pp. 589-603
  4. Libert, J. P., Bach, V. and Farges, G, 1997, 'Neutral Temperature Range in Incubators: Performance of Equipment in Current Use and New Developments,' Critical Reviews in Biomedical Engineering, Vol, 25, No. 4-5, pp. 287-370
  5. Stothers, J. K., 1980, 'The Effect of Forced Convection on Neonatal Heat Loss,' J. Physiol
  6. Stothers, J. K. and Warner, R. M., 1984, 'Thermal Balance and Sleep State in the Newborn,' Early Hum. Dev., Vol, 9, No. 4, pp. 313-322 https://doi.org/10.1016/0378-3782(84)90076-8
  7. Swyer, P. R., 1978, Heat Loss After Birth, in Temperature Regulation and Energy Metabolism in the Newborn, Sinclair, J. C, Ed., Grune and Stratton, New York