A Study on Thermal Diffusivity Measurement by Improvement of Laser Flash Uniformity Using an Optical Fiber

광섬유를 이용한 레이저섬광의 균일분포 증진효과에 따른 열확산계수 측정에 관한 고찰

  • 이원식 (단국대학교 대학원 기계공학과) ;
  • 배신철 (단국대학교 공과대학 기계공학과)
  • Received : 1997.10.28
  • Published : 1998.08.01

Abstract

When thermal diffusivity is measured by laser flash method, the thermal diffusivity call be calculated front the assumption of the uniformly heated whole surface of the specimen. It has been known that the approximate 5% error is made by the non-uniform energy distribution on the specimen surface of laser pulse heat source. In this study, to obtain the highly-uniformed laser beam, which has both the low non-uniform heating error from non-uniform laser beam and the energy loss, research was carried out on no transmitting loss by optical fiber and high repetitions. In addition, heating error and thermal diffusivity were measured as the measuring positions were varied and compared with the results using the uniform and the non-uniform laser beams. In addition, dole to using the uniformalized laser beam, the whole surface of the specimen was heated uniformly and as a result, it was the thought that this was very effective to reduce the variations of the errors of the thermal diffusivity as the measuring positions were varied. It can be obtained that when the thermal diffusivity of POCO-AXM-5Q1 of SRM in NBS was measured with both the uniform and the non-uniform laser beams, the dispersion error of the former was from 2 to 2.5%, which was more improved than that of the latter.