Development and in Vivo Test of an Electrohydraulic Total Artificial Heart at the National Cardiovascular Center in Japan

일본 국립 순환기 센타형 전기유압식 인공심장의 개발과 동물실험

  • 손영상 (고려대학교 의과대학 흉부외과학교실, 일본 국립 순환기병센타 연구소 인공장기부)
  • Published : 1998.04.01

Abstract

The ultimate goal of total artificial heart is permanent substitute for a failed heart in a patient without any other therapeutic modality. Until now, infection has been the main problem related to the mechanical circulatory support system. The best way to solve this catastrophic complication and to improve the quality of life of TAH patients in terms of tethering must be implantation of TAH totally. The EH-TAH has been developed in NCVC from 1987 for this purpose. The system consists of an energy converter and pumps, which are designed to be placed in abdomen and pericardial space separately for a good anatomical fit. To evaluate the anatomical fit and hemodynamic performance of the EH-TAH, in vivo test was done. General condition of the animal and hemodynamic status had been stable until the TAH stopped on the 11th pumping day. The estimated cardiac output was about 7.7L/min. The values of mean aortic pressure, left and right atrial pressure were 93$\pm$10, 19$\pm$3 and 15$\pm$4 mmHg, respectively. The correlation coefficient between left and right atrial pressure was 0.96, which represents the dynamic function of the interatrial shunt in controlling left-right imbalance of cardiac output. During pumping days, the temperature on the surface of actuator had been maintained at 39.7$\pm$0.4$^{\circ}C$, less than 1$^{\circ}C$ higher than the rectal temperature. The TAH stopped on the 11th day due to mechanical problems. We concluded that the EH-TAH possessed satisfactory basic performance including anatomic fit and hemodynamic adequacy, although there were several mechanical problems to be solved yet.

인공심장의 주된 합병증인 혈전과 감염의 문제를 줄이고 장착 환자의 삶의 질을 높이기 위하여 체내 완전 매립형 전기 유압식 인공심장을 개발하고 있는 것과 이 모델을 이용한 동물실험에 대하여 보고하는 바이다. 이 인공심장은 혈액펌프, 유압구동기, 구동제어회로, 경피적 에너지-정보 전송시스템, 체내 받테리 등으로 구성되어 있다. 실험동물은 체중 62Kg의 송아지를 이용하였다. 흉강내 이식하는데 어려움은 없었고, 수술 후 생존하는 동안 혈역학적으로 문제는 없었으며 특이한 합병증도 발생하지 않았다. 그러나 이식 11일째 기계적인 결함으로 작동이 정지하였다. 앞으로 기계적 수명이나 생체적 합성 등의 문제 등히 해결되어야 할 것이다.

Keywords

References

  1. Artificial Organs v.16 no.1 Is a totally implantable artificial heart realistic? Nose Y.
  2. Artificial Organs v.14 no.SUP. 1 Dimensional criteria for the totally implantable artificial heart for human use Taenaka Y;Uyama C;Takano H(et al.)
  3. Artificial Organs v.14 no.SUP. 4 Design and evaluation of a total artificial heart for patients with a smaller chest cavity Taenaka Y;Takano H;Noda H(et al.)
  4. Lab. Animals v.25 Post-operative analgesia following thoractomy in the dog: an evaluation of the effects of bupivacaine intercostal nerve block and nalbuphine on respiratory function Flecknell PA;Kirk AJB;Liles JH;Hayes PH;Dark JH
  5. Am J. Cardiol v.12 Prolonged assisted circulation during and after cardiac or aortic surgery. Prolonged partial left ventricular bypass by means of intracorporeal circulation Liotta D;Hall CW;Walter SH(et al.)
  6. Am J Cardiol v.24 Orthotopic cardiac prosthesis for 2-stage cardiac replacement Cooley DA;Liota D;Hallman RD(et al.)
  7. Assisted Circulation v.3 Toward the totally implantable artificial heart Unger F.
  8. Proc Eur Soc Artif Organs v.6 Long-ter results with seamless blood pumps out of polyurethanes for the replacement of the heart Rennekamp F;Lemm W;Zartnack F(et al.)
  9. Artificial Organs v.19 no.6 Constructional and functional characteristics of recent total aritificial heart models TNS Brno VII, VIII and IX Vasku J;Urbanek P
  10. Artificial Organs v.10 no.3 Thromboembolic complications of the Jarvik-7 total artificial heart: case report Levinson MM;Smith RG;Cork RC(et al.)
  11. Trans Am Soc Artif Intern Organs v.33 no.3 Early preliminary results of a new total artificial heart connector system Holfert JH;Riebman JB;Dew PA(et al.)
  12. Am Soc Artif Intern Organs v.38 no.3 An electrically powered total artificial heart. Over 1 year survival in the calf Synder A;Rosenberg G;Reibson J(et al.)
  13. ASAIO J v.39 An atrial hydraulic shunt in a total artificial heart. A balance mechanism for the bronchial shunt Kung RTV;Yu LS;Ochs B;Parnis S;Frazier OH
  14. Artificial Organs v.12 In vitro development of automatic control for the activity filled electrohydraulic heart Lioi AP;Orth JR;Crump KR(et al.)
  15. Artificial Organs v.17 no.3 Left and right pump output control in one-piece electromechanical total artificial heart Takatani S;Shiono M;Sasaki T(et al.)
  16. Int J Artif Organs v.16 no.1 Different stroke volumes for the left and right ventricles in the moving-actuator type total artificial heart Min B;Kim I;Kim H
  17. Int J Artif Organs v.14 no.6 Right-left ventricular output balance in the totally implantable artificial heart Olsen DB;White RK;Long JW;Khanwilkar PS
  18. Am Soc Artif Intern Organs v.38 no.3 A blood pump with an interatrial shunt for use as an electrohydraulic total artificial heart Tatsumi E;Diegel PD;Holfert JW(et al.)
  19. Am Soc Artif Intern Organs v.37 no.3 Determination of atrial shunt size needed to balance an electrohydraulic total artificial heart Kinoshita M;Hansen CA;Khanwilkar PS;Whith KR;Olsen DB