Several types of electrical artifact seen on electroencephalogram( EEG) records are described. Those are the EOG and the PVC roller pump noise, and so on. An adaptive digital filtering of the electroencephalogram( EEG) is a successful way of suppressing mains interference, but it affects some of the frequency components of the signal, whore artifacts may not be acceptable in some cafes of automatic EEG processing. Thus we studied the method for cancelling these artifacts. This proposed method does not use the reference channel, and is realized by connecting the linear predictor and the fixed FIR filter for the EOG artifact, and by cascading the linear predictor and the noise canceller for the pump artifact. The simulation results illustrate the performances of the proposed method in terms of the capability of interferences suppression. In the results we obtained about 20 dB noise reduction.

본 논문에서는 뇌파측정기에서 발생하는 전기적 잡음을 제거하기 위하여 뇌파 측정시 발생하는 안구운동, PVC 룰러 펌프 잡음등을 포함하는 전기적 잡음에 대하여 기술하였다. 적응 디지틀 필터는 이러한 뇌파 측정 잡음을 제거하는 효과적인 방법이다. 그러나 일반적인 적응 디지틀 필터는 뇌파신호의 일부주파수 성분에 영향을 주게되므로 자동 뇌파 처리에는 부적합하다. 그러므로 본 고에서 이러한 뇌파잡음을 제거하기 위하여 제안한 방법은 일반적인 적응디지틀 필터의 참조채널을 사용하지 않으며, EOG 잡음을 제거하기 위하여 선형예측기와 고정차수 FIR 필터를 연결하였으며, 펌프잡음을 제거하기 위하여 선형예측기와 잡음 제거기를 연결함으로써 실현하였다. 또한 시뮬레이션을 통한 결과는 잡음제거 능력의 관점에서 제안한 방법의 성능을 분석하였으며, 결과적으로 약 20dB의 잡음제거 성능을 보였다.

Keywords

References

  1. Electroencephalography:Basic principles, Clinical Applications, and Related Fields,(3rd ed.) E. Niedermeyer;F. L. Da Silva
  2. J. of Clinical Eng. v.20 no.1 Artifact induced by roller pump tubing on EEG tracing of neonatal ECMO patients E.J. Snyder;G.J. Shovlin;S.G. Beacham;J. Gensink
  3. IEEE Trans. Biomedical Eng. v.BME-30 no.7 ECG enhancement by adaptive concellation of electrosurgical interference M. Yelderman;B. Widrow;J. M. Cioffi;E. Hesler;J. A. Leddy
  4. IEEE Trans. Biomedical Eng. v.BME-29 Fetal electrocardiogram enhancement by time-sequenced adaptive filtering E.R. Ferrara;B. Widrow
  5. IEE Proc. v.137 no.1 Detection and suppression of impulsive noise in speech communication system S.V. Vaseghi;P.J.W. Rayner
  6. J. of Clinical Neuropnsiology v.2 no.3 Methods of analysis of nonstationary EEGs with emphasis on segmentation techniquesic comparative review J.S. Barlow
  7. J. of Clinical. Eng. v.20 no.2 A neuadaptive cohernet model algorithm for removal of power-line interference L. Gang;Y. Qilian;Y. Xuemin
  8. Digital, signal processing-a practival approach E.C. Ifeachor;B.W. Jervis