The Characteristics of Two-dimensional Radiation Monitoring System

2차원 방사선준위 분포측정 시스템의 제작 및 특성 조사

  • Kim, S.H. (Dept. of Physics, Kyungpook National University) ;
  • Han, S.H. (Dept. of Physics, Kyungpook National University) ;
  • Kim, K.R. (Dept. of Physics, Kyungpook National University) ;
  • Kang, H.D. (Dept. of Physics, Kyungpook National University) ;
  • Choi, Y.S. (Korea Advanced Energy Research Institute) ;
  • Lee, Y.B. (Korea Advanced Energy Research Institute) ;
  • Lee, J.M. (Korea Advanced Energy Research Institute)
  • Published : 1997.07.31

Abstract

In this study, the radiation monitoring system using CsI(TI) scintillation counter is developed for the measurement of radiation distribution in the field of high dose level. When the inner diameter of collimator is 8 mm, we have realized the optimum detecting efficiency and spatial resolution. At that time, the position resolution was 10 cm at 1 m from the system. And experimental results indicated that the energy resolution of the system were 10 % for 662 keV of Cs-137, 7.6 % for 1.17 MeV of Co-60, and 5.8 % for 1.33 MeV of Co-60. Also, we have shown that the real radiation distribution images may be obtained by our measurement system.

본 연구에서는 CsI(TI) 섬광계수기를 이용하여 고방사선장에서 방사선준위분포를 측정할 수 있는 2차원 측정시스템을 개발하였다. 콜리메이터의 내경이 8 mm일 때 최적의 위치분해능과 검출효율을 얻었으며 이때 위치분해능은 1 m 거리에서 10 cm미만이었다. 그리고 시스템의 에너지 분해능은 Cs-137의 662 keV에 대하여 10 %, Co-60의 1.17 MeV에 대하여 7.6 %, 1.33 MeV에 대하여 5.8 %를 얻었다. 그리고 실제 방사선 선원에 대하여 그 분포를 측정할 수 있었다.

Keywords