Measurement of Detection Efficiency of Plastic Track Detector to ${\alpha}-particle$
고체비적검출기의 알파선 검출효율 측정에 관한 연구
- Park, Young-Woong (Korea Atomic Energy Research Institute) ;
-
Yang, Hyun-Soo
(Chungnam National University)
- Published : 1996.09.30
Abstract
The calibration method for the detection efficiency of solid state track detector to alpha is presented, and the detection efficiency of CN-85 and LEXAN to alpha is measured by this method. The results are 97% and 57% in CN-85 and LEXAN, respectively. The reason for a better result in CN-85 in not only the high sensitivity but also the low energy dependency with comparing for LEXAN. The peak detection efficiency is 3MeV and 1.8MeV in CN-85 and LEXAN, respectively.
알파선 검출 효율 측정을 위한 고체비적검출기의 교정 방법을 제시하였으며, 제시된 교정 방법을 이용하여 CN-85와 LEXAN의 검출 효율을 측정하였다.
Keywords