A Study of X-ray Output for Diagnostic X-ray Equipment

진단용 X선 발생 장치의 X선 출력에 관한 연구

  • Ko, Shin-Gwan (Dept. of Radiotechnology, Dae Jon Health Junior College) ;
  • Ahn, Bong-Seon (Dept. of Radiotechnology, Dae Jon Health Junior College) ;
  • Jang, Sang-Sup (Dept. of Radiotechnology, Dae Jon Health Junior College) ;
  • Choi, Jong-Woon (Dept. of Radiotechnology, Dae Jon Health Junior College) ;
  • Shin, Young-Soon (Dept. of Radiotechnology, Dae Jon Health Junior College)
  • 고신관 (대전보건전문대학 방사선과) ;
  • 안봉선 (대전보건전문대학 방사선과) ;
  • 장상섭 (대전보건전문대학 방사선과) ;
  • 최종운 (대전보건전문대학 방사선과) ;
  • 신영순 (대전보건전문대학 방사선과)
  • Published : 1995.12.30

Abstract

For the managements of the diagnostic X-ray equipments, the authors examined the output of single phase rectification assembly, Three phase rectification assembly and serial radiographic appartus, and got the following conclusions. 1. When the tube voltages in X-ray control panels ware compared to the measured values on the kVp pulse meter, only little differences were detected in all the X-ray equipments. And most of the equipments were all well managed within the internationally permitted limits, excepting the 12.02 % error at 120 kVp in three phase rectifying assembly. 2. As for the X-ray qualities affecting the X-ray images, the serial radiographic apparatus showed excellence, while the single phase rectification assembly were somewhat inferior to the others only maining the internationally recommended limits. 3. The tube voltage ranges where the X-ray output showed excellence were $100{\sim}200\;mA$ in serial radiographic apparatus, $200{\sim}350\;mA$ in three phase rectification assembly and $350{\sim}400\;mA$ in single phase rectification assembly respectively. 4. In the repeatability test of the X-ray equipments, CVs were in the range of $0.0029{\sim}0.049$, which is within the HEW or KS standards. Consequently all the equipments are thought to be well-manage. 5. This study on characteristics and output of the X-ray equipments was accomplished within a limited short time. Long-time researches on the function managements for the X-ray equipments should be followed along with the periodical checking the output for reduction of X-ray exposures to the patients or radio-technologists, and for maintanance and prediction of trouble of the equipments.

Keywords