Journal of radiological science and technology (대한방사선기술학회지:방사선기술과학)
- Volume 18 Issue 2
- /
- Pages.61-73
- /
- 1995
- /
- 2288-3509(pISSN)
- /
- 2384-1168(eISSN)
A Study of X-ray Output for Diagnostic X-ray Equipment
진단용 X선 발생 장치의 X선 출력에 관한 연구
- Ko, Shin-Gwan (Dept. of Radiotechnology, Dae Jon Health Junior College) ;
- Ahn, Bong-Seon (Dept. of Radiotechnology, Dae Jon Health Junior College) ;
- Jang, Sang-Sup (Dept. of Radiotechnology, Dae Jon Health Junior College) ;
- Choi, Jong-Woon (Dept. of Radiotechnology, Dae Jon Health Junior College) ;
- Shin, Young-Soon (Dept. of Radiotechnology, Dae Jon Health Junior College)
- 고신관 (대전보건전문대학 방사선과) ;
- 안봉선 (대전보건전문대학 방사선과) ;
- 장상섭 (대전보건전문대학 방사선과) ;
- 최종운 (대전보건전문대학 방사선과) ;
- 신영순 (대전보건전문대학 방사선과)
- Published : 1995.12.30
Abstract
For the managements of the diagnostic X-ray equipments, the authors examined the output of single phase rectification assembly, Three phase rectification assembly and serial radiographic appartus, and got the following conclusions. 1. When the tube voltages in X-ray control panels ware compared to the measured values on the kVp pulse meter, only little differences were detected in all the X-ray equipments. And most of the equipments were all well managed within the internationally permitted limits, excepting the 12.02 % error at 120 kVp in three phase rectifying assembly. 2. As for the X-ray qualities affecting the X-ray images, the serial radiographic apparatus showed excellence, while the single phase rectification assembly were somewhat inferior to the others only maining the internationally recommended limits. 3. The tube voltage ranges where the X-ray output showed excellence were
Keywords