Design and Filtration Performance of Polysulfone Hollow Fiber Membrane Hemofilter

폴리설폰 중공사막 혈액여과기의 설계 및 여과성능

  • 김재진 (한국과학기술연구원 분리막연구실) ;
  • 박진용 (한국과학기술연구원 분리막연구실, 녹십자의료공업(주) 중앙연구소)
  • Published : 1993.09.01

Abstract

Hemofilter was optimally designed and manufactured using polysulfone hollow fiber mem- brane with surface area of 0.6mE Molecular weight cut-off of the hemofilter was measured with polyethylene glycol and dextran aqueous solutions of various molecular weights and it was ranged from 9,500 to 38,900. Ultrafiltration rates were measured with pure water in a static system and flowing system. The clearance of urea, creatinlne, and vitamine BIB were measured using aqueous solutions.

Keywords

References

  1. Trans. Am. Soc. Artif. Intern. Organs v.16 Uremic Blood Cleansing by Diafiltration Using a Hollow Fiber Ultrafiler L.W.Henderson;C.Ford;C.K.Colton;L.W.Bluemle;H.J.Bixler
  2. Proc. Eur. Dialy. Transplant Assoc. v.13 Treatment of Chronic Uremia by an Ultrafiltration Kidney E.Quellhorst;J.Rieger;B.Doht;H.Beckemann;I.Jacob;B.Kraft;G.Mietzch;F.Scheller
  3. 賢と透析 v.3 大田和夫;鈴木利昭;小作保雄;佐中孜;星野敏久;阿岸鐵三;杉野信傳
  4. Hemofiltration L.W.Henderson;E.A.Quellhorst;C.A.Baldamus;M.J.Lysaght
  5. Trans. Am. Soc. Artif. Intern. Organs v.17 Blood Cleansing by Diafiltration in Uremic Dog and Man R.Henderson;C.Ford;C.Colton;R.Cross;S.Steinmuller;L.Henderson
  6. 化學增刊 v.92 丹澤宏;酒非良忠
  7. Trans. Am. Soc. Artif. Intern. Organs v.28 Hemodiafiltration Using a Hybrid Membrane System for Self-generation of Diluting Fluid A.C.Cheung;Y.Kato;J.K.Leypolt;L.W.Henderson
  8. 纖維素系樹脂 金傳允
  9. 한국과학기술연구원 연구보고서 첨단의료기술 개발 : 인공신장 여과기 개발 김재진(외)
  10. 한국과학기술연구원 연구보고서 한외여과용 고분자막 제조(1), (2), (3) 김재진(외)