The Experiment of Grid Characteristics for High-voltage Radiography of Chest

흉부촬영시 관전압과 선질에 따른 적절한 격자의 선택을 위한 실험

  • Kim, Jung-Min (Dept. of Radiotechnology, Ji San Junior College) ;
  • Ahn, Bong-Seon (Dept. of Radiotechnology, Dae Jon Health Junior College)
  • 김정민 (지산간호보건전문대학 방사선과) ;
  • 안봉선 (대전보건전문대학 방사선과)
  • Published : 1992.12.30

Abstract

Grids can improve the diagnostic quality of chest radiography by trapping the greater part of scattered radiation thus providing more detailed. chest radiographic images. It is most effective mathod of reduce the scatter ratio but must increase the expour factor. The benefit of use of grid is improve the contrast and the loss is increase of patient dose. In chest radiography especially hard quality high voltage radiography it will have to be considered to select the optimum grid with view point of benefit and loss. In this experiment, author got some result of characteristics about 4 different grids with film method. 1. There was no difference the scatter ratio in case of no grid and the scatter ratio was about 60%. 2. 16 : 1 grid was excellent of scatter reduction factor in high voltage chest radiography, next was 10 : 1, CROSS, MICRO FINE grid have low scatter reduction rate compare to 16:1, 10:1 grid. 3. The bucky factor of CROSS grid in accordance of kVp was find out the highest in 4 grids, on the contraly 10 : 1 grid was profitable to the exposure does. 4. With careful consideration in the point of scatter reducion rate and bucky factor, author suggest the 10 : 1 linear grid on the use of chest radiography in $80{\sim}120\;kVp$, 16 : 1 grid in $120{\sim}140\;kVp$.