High Energy Photon Beam Modeling Using Transport Theory for Calculation of Absorbed Dose Distribution

흡수 선량 분포의 수송방정식을 이용한 10 MV X-선의 모델

  • Choi, Dong-Rak (Department of Radiology, Catholic University Medical College) ;
  • Chun, Ha-Chung (Department of Therapeutic Radiology, Hanyang University Hospital,) ;
  • Lee, Myung-Za (Department of Therapeutic Radiology, Hanyang University Hospital,)
  • 최동락 (가톨릭대학교 의과대학 방사선과학교실) ;
  • 전하정 (한양대학교 의과대학 치료방사선과학교실) ;
  • 이명자 (한양대학교 의과대학 치료방사선과학교실)
  • Published : 1992.06.01

Abstract

A mathematical model is presented for the calculation of the depth absorbed dose in water Phantom irradiated by high energy Photon beam (10MV X-ray), based on transport theory. The parameters of this model are obtained from the experimental values which were simulated by non-linear regression process method. The calculated absorbed dose distribution is extended to 3-D by using trial function from beam profile field sizes, SSD and depth in water phantom irradiated by high energy Photon beam. The calculated values using this model are in good agreement with the measured values.

물팬톰내에 조사된 10 MV X-선의 심부율을 입자의 수송이론을 근거로 한 1차원적인 모델을 이용하여 계산하였다. 계산된 이론식의 매개상수는 9개로 줄일 수 있었으며 실측치를 이용하여 비선형 회귀분석 방법으로 얻을 수 있다. 조사면과 선원간의 거리 및 깊이에따른 3차원적인 흡수선량분포의 계산식은 고에너지 광자선이 조사된 물팬톰내에서의 Beam Profile에 대한 시도함수를 이용하여 수송이론에의한 심부율계산을 3차원적으로 확장하였으며 흡수 선량 분포는 3차원적 위치의 함수로 널리 계산할 수 있다. 이 모델을 사용하여 계산된 이론값은 실험값과 $\pm12\%$ 이내의 만족할만큼 잘 일치하였다.