Evaluation of Automatic Exposure Control Systems
X선자동노출(X線自動露出) 제어장치(制御裝置)에 관한 검토(檢討)
- Kim, Chung-Min (Dept. of Radiotechnology, Jisan Junior College) ;
- Kim, Keon-Chung (Dept. of Diagnostic Radiology, Asan Medical Center) ;
- Huh, Joon (Dept. of Radiotechnology, Junior College of Allied Health Sciences)
- Published : 1991.05.25
Abstract
Automatic exposure control systems are now readily available and widly used in radiography. A device that automatically determines the amount of radiation required to produce an acceptable level of film blackness is called a phototimer, or automatic exposure control. There are limitations in the use of phototimers ; however, equipment can be used much more effectively if one recognize these limitations, and the goal of obtaining a properly exposed, reproducible radiography can be achieved. We have some results in this experimental study. The variations of film density due to kVp are in propotion to below 80 kVp range. But indicate constant level above 80 kVp to 120 kVp. The characteristic density variation by phantom thickness shows the higher the thiner in exit type. AEC system have a density control that is designed to make adjustments increase or decreas the density 10% per step or 25% per step. The automatic exposure control circuits must meet the several regulations of the HEW. Some forms of technique chart is recomened for use of AEC.