A Study on the Radiation Induced Current of X-ray Insulating Oil

X선 장치용 절연유의 방사선 유기전류에 관한 연구

  • 김영일 (신구전문대 방사선과) ;
  • 정연택 (명지대 공대 전기공학과) ;
  • 이덕출 (인하대 공대 전기공학과)
  • Published : 1989.07.01

Abstract

Radiation induced current of the insulating oil used for X-ray tube housing was investigated in this paper. The X-ray induced current increased in proportion to the X-ray intensity and X-ray wave length and was greatly affected by the temperature of the insulating oil and the gap length` however, the current was not varied by the amount of the X-ray absorbed. The results suggest that one must consider the following problems as important factors for X-ray machine design and normal operation: cooling system of the X-ray tube housing, X-ray intensity, increasing temperature due to continuous operation, years used, and oil exchange and management.

Keywords