Journal of radiological science and technology (대한방사선기술학회지:방사선기술과학)
- Volume 7 Issue 1
- /
- Pages.47-52
- /
- 1984
- /
- 2288-3509(pISSN)
- /
- 2384-1168(eISSN)
A Study of Total Filtration of the Diagnostic X-ray Equipments
진단용(診斷用) X선장치(線裝置)의 총여과(總濾過)에 관(關)한 조사연구(調査硏究)
- Kim, Young-Il (Dept. of Radiotechnology, Shin Gu Junior College) ;
- Kyong, Kwang-Hyon (Dept. of Radiation Standards, National Institute of Health Republic of Korea)
- Published : 1984.12.15
Abstract
This paper was performed research for the total filtration in the useful beam of 45 diagnostic x-ray equipments using in Korea. We obtained the following results: 1. A good ability of total filtration was showed in higher than lower maximum permissible rating equipments. 2. In 53 percentage of 45 diagnostic x-ray equipments, the total filtration in the useful x-ray beam was not less than the values given in the Korean industrial standard. This results suggested to user to need the more ability management of diagnostic x-ray equipments using now.