Study on Sensitivity and Resolution of Multihole Focusing Type Collimator

다공초점형(多孔焦点型) 콜리메타의 감도(感度)와 해상력(解像力)에 관(關)한 검토(檢討)

  • Park, Sung-Ock (Dept. of Radiotechnology, Dong Nam Junior Health College) ;
  • Kyong, Kwang-Hyon (Dept. of Radiotechnology, Shin Heung Junior College) ;
  • Park, Jun-Chul (Dept. of Radiotechnology, Incheon Jonior Nursing College and Arrange Health Sciences)
  • 박성옥 (동남보건전문대학 방사선과) ;
  • 경광현 (신흥실업전문대학 방사선과) ;
  • 박준철 (인천간호보건전문대학 방사선과)
  • Published : 1982.12.15

Abstract

In order to estimate characteristics and functions of multi hole focusing type collimator, the authors measured the sensitivity and resolution varing with the distance between the face of collimator and source by experimental study. The results obtained with this study are as follows. 1. In comparison with focal distance obtained fram sensitivity measurement of various multi hole focusing type collimator, collimator A, C, E was 10 cm and collimator B, D was 12 cm in focal distance. 2. In regard of resolution of collimators having 10 cm focal distance, collimator A showed the finest resolutions than any other collimator(C, E). 3. By comparison regarding sensitivity by focal dist ante, the shorter the distance than a give focal distance of collimator, the worse the sensitivity of collimator. 4. With regard to resolution by focal distance, the shorter the distance than a given focal distance of collimator, the worse the resolution of collimator.

Keywords