Detection of Plethysmogram and its Usefulness

맥파검출과 이의 유효성

  • 홍승홍 (인하대학교 공과대학 전자공학과)
  • Published : 1978.03.01

Abstract

This paper described the establishment of the basic data for the design of the transducer for the detection of plethysmogram, the structure of pulsimeter utilizing the designed transducer and the clinical usefulness of detected waveform of the optical detection method can be utilized as an information of cardiovascular system as physiological theoretical waveform of aorta, and derivative of plethysmogram by piezo electric elements clearly shows the information of incisura which does not appear at the old age and this makes the measuring of ejection time to be easy. The ejection time of average for 42 persons shows 0.324 $\pm$ 0.027 sec. and this value is in good agreement with the theoretical value, and the possibility of measuring the velocity of pulse conduction has been studied.

맥파검출용변환기설계의 기초자료 확립과 설계된 변환기를 이용한 맥추계구성 및 검출된 파형의 임상적 유효성을 검토했다. 광학적 검출방식에 의한 출력파형은 대동장의 생리학적인 이론파형과 동일한 파형으로 맥관계의 한 정보로써 이용할 수 있으며 여전소자에 의한 맥파의 미분파형은 고령자에게서 잘 나타나지 않는 절역의 정보가 분명히 표시되어져 임혈시간의 측정이 용이하다. 42명에 대한 평균임혈시간이 0.324±0.027초로 얻어져 생리학적 이론치와 같으며 맥파전달시간의 측정가능성이 검토되어졌다.