A prototype of photoelectric photometer has been developed by all solid state operational amplifiers. The amplification of low-level anode currents from photomultiplier tube corresponding to a given incident light intensity and logarithmic transfer for star's magnitude nleasurements are achieved by the same operational amplifier. It is important to select low input bias currents for stable performance because the resolution is primarily limited by the bias currents. As such, they offer a higher reliable and economical for a large number of electrometer amplifier applications which have traditionaly been fulfilled by vacuum electrometer tubes or vibrating reed electrometers.
고체화된 연산증폭기를 중심으로 구성한 광전측광장치를 구성하였다. 광원에 대응하여 광전자증배관에서 얻는 미소전류의 증폭과 별의 등급측정을 위한 대수변환은 동일연산증폭기의 변형으로 얻고 있다. 측정기능의 한계는 주로 증폭기입력단의 bias전류에 좌우되므로 이것은 광산증폭기의 선택상 주요기준이 될 것이다. 이러한 연산증폭기는 종래 전위계진공관 혹은 진동용량형전위계 등으로 가능하였던 각종 미소전류·전압의 증폭에 높은 신뢰성과 경제적인 증폭방식으로 적용된다.