Phantom of the AAPM CT imaging evaluation Studies on the quantitative analysis method

CT 정도관리 영상의 정량적 분석방법에 관한 연구

  • Kim, Young-su (Department of Radiology, Daewoo General Hospital) ;
  • Ko, Seong-Jin (College of Health Sciences, Catholic University of Pusan) ;
  • Kang, Se-Sik (College of Health Sciences, Catholic University of Pusan) ;
  • Ye, Soo-young (College of Health Sciences, Catholic University of Pusan)
  • 김영수 (대우병원 영상의학과) ;
  • 고성진 (부산 가톨릭 대학교 보건과학대학 방사선학과) ;
  • 강세식 (부산 가톨릭 대학교 보건과학대학 방사선학과) ;
  • 예수영 (부산 가톨릭 대학교 보건과학대학 방사선학과)
  • Published : 2016.05.25

Abstract

CT quality assurance imaging evaluation and enforcement as quantitative assessment by phantom image evaluation, assessment items include There are also contrasting the water attenuation coefficient, uniformity, noise, resolution, spatial resolution, 10mm slice thickness evaluation, contrast resolution, space for the resolution, the slice thickness evaluation, it is possible to estimate the error due to the evaluation by the subjective judgment of the tester, using a subjective error image processing program to be computed to minimize the objective evaluation. Basic recording conditions of the CT image quality control assessment is the same as special medical equipment quality control checks, the images were evaluated quantitatively using IMAGE J. For a CT attenuation coefficient, the uniformity, noise evaluation, were evaluated as CT quality control image the standard deviation of the measured value of the digital processing of image smaller and less noise uniform images than the, contrast and resolution assessment is the size of the diameter of a circle having a large the 1 inch, 0.75 inch, 0.5 inch quality if the diameter of the circle, was evaluated in the small circle in the near circle ellipse. Spatial resolution is evaluated by using a self-extracting features of an image processing program, all of the groups of members comprising the acceptance criteria to automatically extract, was evaluated to be very useful for the quantitative assessment. When CT image quality control assessment on the basis of the results such as the above, if using an image processing program to minimize the subjective judgment of the error evaluator and is determined more efficient than would be made quantitative evaluation.

CT 정도관리 영상평가는 팬텀 영상평가를 통한 정량적인 평가로 시행하며, 평가 항목으로는 물의 감약 계수, 균일도, 노이즈, 대조도 분해능, 공간분해능, 10mm 슬라이스 두께 평가 등이 있으며, 대조도 분해능, 공간 분해능, 슬라이스 두께 평가의 경우, 검사자의 주관적인 판단에 의한 평가로 인한 오류를 예상할 수 있으므로, 주관적인 오류를 최소화 하고자 전산화된 영상처리 프로그램을 이용하여 객관적인 평가를 하고자 한다. CT 정도관리 영상평가의 기본 촬영 조건은 특수의료장비 품질관리검사와 동일하며, IMAGE J를 이용하여 영상을 정량적으로 평가하였다. CT 감약계수, 균일도, 노이즈 평가의 경우, CT 정도 관리 영상에 비하여 디지털 처리 영상의 측정값의 표준편차가 더 작아 잡음이 적고 균일한 영상이라고 평가하였으며, 대조도 분해능 평가는 원의 직경의 크기가 큰 1인치, 0.75인치, 0.5인치의 경우 원형, 원의 직경이 작아질수록 타원에 가까운 원형으로 평가되었다. 공간분해능 평가는 영상처리 프로그램의 자동추출 기능을 이용하여, 합격기준을 포함하는 원의 그룹을 모두 자동으로 추출하여, 정량적인 평가에 매우 유용하다고 평가하였다. 위의 결과 등을 바탕으로 CT 정도관리 영상 평가 시, 영상처리 프로그램을 이용한다면 평가자의 주관적인 판단 오류를 최소화하고, 보다 더 효율적인 정량적평가가 이루어 질 것이라고 판단된다.

Keywords