Setup and Atomic Calibration of Particle Induced X-ray Emission System

  • Published : 2014.02.10

Abstract

Recently, particle induced X-ray emission (PIXE) analysis system was installed at the 2MV ion acceleration system in Korea institute science and technology (KIST). This installation is for complement to low atomic resolution of heavy atoms at Rutherford backscattering spectrometry (RBS) system. For quantitative analysis, a mass calibration of the PIXE set-up has been done with thin film standards and. The GUPIX software package has been used to process the PIXE spectra and the results are compared with the values from RBS system. Therefore, the instrumental constant H (solid angle and correction factor) is determined relying completely on the GUPIX data base (cross-sections, fluorescence and Coster-Kronig probabilities, stopping powers and attenuation coefficients) for a large set of elements. These H values can be used in future analysis.