Computed Radiography applied to the Department of Radiation Oncology

Computed Radiography의 방사선종양학과로의 적용

  • Hong, Seung-Il (Dept. of Radiological Science Technology Gwangyang Health College) ;
  • Song, Jong-Nam (Dept. of Computer Science Sunchon National University) ;
  • Kim, Young-Jae (Dept. of Public Health and Medicine Dongsin National University)
  • Published : 2011.10.26

Abstract

We measured that is Gantry, Collimator Star Shot, Light vs. Radiation, HDR QA with Medical LINAC Then, PACS was implemented on the digital images on the monitor that can be confirmed through the QA. Also, for cooperation with OCS system that is using from present source and impose code that need in treatment in each treatment, did so that Order that connect to network, input to CR may appear, did so that can solve support data mistake of Pinacle and PACS that is Planning System and look at Planning premier in PACS.

본 논문에서는 의료용 선형가속기를 이용하여 현재 시행하고 있는 정도관리 중 Gantry, Collimator Star Shot, Light vs. Radiation Field Accuracy, HDR QA(Dwell position accuracy)를 시행하여 PACS 상에 구현하였고 모니터 상에서 디지털 영상을 통한 QA가 가능한지 확인하였다. 또한, 현재 본원에서 사용 중인 OCS system과 연동하여 치료에 필요한 코드를 각각의 치료에 부과하여 네트워크로 연결, CR상에 입력한 Order가 나타나도록 하였으며, Planning System인 Pinacle과 PACS상의 지원 data 오류를 해결하여 PACS 상에서도 Planning 영상을 볼 수 있도록 하였다.

Keywords