Thermopile Radiometer Calibration Using Reference Instrument

표준준기에 의한 일사계 교정

  • Published : 2010.06.17

Abstract

The main purpose of the calibration procedure is to perform a one to one comparison of the reference pyranometer and the test pyranometer. In order to achieve this, both pyranometers need to be exposed to exactly the same irradiance, under the same circumstances. There are a number of error sources that could result in a wrong measurement. Most importantly Lamp instability, pyranometer offsets, thermal offsets of junctions, voltmeter offset, voltmeter instability, reference pyranometer instability, tilting of the pyranometers and differences in sensor height. Another sun-disc calibration procedure compares the computed vertical component of the direct irradiance as measured by a pyranometer with that measured by the pyranometer to be calibrated. Readings are taken with the levelled pyranometer on a clear day. Firstly the global irradiance and then the diffuse component are measured. Simultaneously measurement of direct irradiance is made with the pyrheliometer. The ways of performing the calibration and the subsequent calculation have been chosen such that the effect all these error sources has been eliminated as much as possible.