초록
To diagnose vascular diseases from the viewpoint of hemodynamics, detailed quantitative hemodynamic information of related blood flows with a high spatial resolution of tens micrometer and a high temporal resolution in the order of millisecond were required. For investigating the hemodynamic phenomena of vascular circulatory diseases, a new diagnosing technique combining a medical radiography and PIV method was newly developed. This technique consisted of a medical X-ray tube, an X-ray CCD camera, a shutter module for double pulse-type X-ray source, and a synchronizer. Through several preliminary tests, the feasibility of the angiographic PIV technique was verified. For in-vivo measurements to real blood flows, we developed tracer microcapsules which were optimized to this system by encapsulating a contrast material of iodine with a matrix material of PVA (polyvinylpyrrolidone). In near future, the angiographic PIV technique would be useful for diagnosing hemodynamic phenomena of vascular diseases and for their early detection.