CCD image system을 갖는 유전체칩 분석장치를 위한 효율적인 광학시스템 설계

Effective Optical System Design for Biochip Analyzer with CCD Image System

  • 배수진 (한국전기연구원 전자의료기기연구그룹) ;
  • 강욱 (한국전기연구원 전자의료기기연구그룹)
  • 발행 : 2003.11.21

초록

Biochip-based diagnostic technology is an effective, time- and money- saving way. But, biochip analyzer including CCD imaging system has a complete optical system. It is one of reasons that the cost of biochip analyzer is expensive with CCD imaging system. In this paper, We suggested the simple and effective optical system for biochip analyzer with CCD imaging system. It consists of two parts with the same structure but opposite direction. Each part consists of achromatic doublet and meniscus. Suggested optical system has less lenses than existing system and more efficiently.

키워드