Measurement of skin moisture using a FT-NIR spectrometer

  • Suh, Eun-Jung (College of Pharmacy, Dongduk Women's University) ;
  • Woo, Young-Ah (College of Pharmacy, Dongduk Women's University) ;
  • Kim, Hyo-Jin (College of Pharmacy, Dongduk Women's University)
  • Published : 2003.10.01

Abstract

In this study, a FT-NIR spectroscopy was used to determine skin moisture. NIR diffuse reflectance spectra were collected from separated dorsal and abdominal hairless mouse skin. Partial least squares regression (PLSR) was applied to develop calibration models that determine the water content. The seven spectra regions, such as 833-2500, 1100-2250, 1100-1750, 1750-2250, 1200-1600, 1800-2200, and 1200-2200 except 1600-1800 nm, were investigated for the best model by PLSR. The developed model predicted skin moisture for validation set with a standard errors of prediction (SEP) of 4.43%, when used 833-2500 nm. (omitted)

Keywords