S-ketoprofen enantiomer의 정제를 위한 SMB(Simulated Moving Bed) 크로마토그래피 장치의 모델링
- Published : 2002.04.12
Abstract
We have performed batch chromatography experiments to calculate parameters which are used to design SMB chromatography system. Isotherms of S- and R-ketoprofen enantiomers were gained from small amount loading experiments in a column, and flow rate of four zones of SMB chromatography were calculated using mass balance equations. As a results, diameter and length of columns were 10mm and 600mm, and flow rate of each zone were as follows;
Keywords