Optimization of Data Acquisition System with Parallel Collection for PET

  • Yoshida, Eiji (National Institute of Radiological Sciences, Chiba, Japan) ;
  • Shimizu, Keiji (Hamamatsu Photonics K.K., Shizuoka, Japan) ;
  • Murayama, Hideo (National Institute of Radiological Sciences, Chiba, Japan)
  • Published : 2002.09.01

Abstract

We are under development of a 3D PET scanner with depth of interaction (DOI) capable of high sensitivity and high resolution. In this scanner, a maximum data transfer rate of coincidence pair's event information is 10 Mcps and one event is a 64-bit data format. This maximum data transfer rate corresponds by 10 times a conventional PET scanner. A data acquisition system, which fulfills the specification of this scanner, is considered for parallel collection with banks including several coincidence units. Data transfer rate is improved by optimizing parameters of a message size, and so on.

Keywords