Line Image Correction of the Positron Camera in the Secondary Beam Course of HIMAC
- Iseki, Yasushi (Dept. of Energy Sciences, Tokyo Institute of Technology, Nagatsuta, Yokohama, Japan, Power & Industrial Systems R&D Center, Toshiba Corp.) ;
- Mizuno, Hideyuki (Saitama Cancer Center, Ina-machi, Saitama, Japan) ;
- Kanai, Tatsuaki (National Institute of Radiological Sciences) ;
- Kanazawa, Mitsutaka (National Institute of Radiological Sciences, Dept. of Energy Sciences, Tokyo Institute of Technology, Nagatsuta, Yokohama, Japan) ;
- Kitagawa, Atsushi (National Institute of Radiological Sciences) ;
- Suda, Mitsuru (National Institute of Radiological Sciences) ;
- Tomitani, Takehiro (National Institute of Radiological Sciences) ;
- Urakabe, Eriko (National Institute of Radiological Sciences)
- Published : 2002.09.01
Abstract
A positron camera, consisting of a pair of Anger-type scintillation detectors, has been developed for verifying the ranges of irradiation beams in heavy-ion radiotherapy. Images obtained by a centroid calculation of photomultiplier outputs exhibit a distortion near the edge of the crystal plane in an Anger-type scintillation detector. The images of a