Development of a neutron Dosimeter using PIN diode

핀(PIN) 다이오드 소자를 이용한 중성자 측정장치 개발

  • Published : 2001.07.18

Abstract

Si PIN diodes are subject to be damaged from the exposure of fast neutron by displacement of Si lattice structure. The defects are effective recombination centers for carriers which migrate through the base region of the PIN diode when forward voltage is applied. It causes an increase in current and a decrease in resistivity of the diode. This paper presents the development of a neutron sensor based on displacement damage effect. PIN diodes having various structures were made bymicro-fabrication process, and neutron beam test was performed to identify neutron damage effect to the diode. From a result of the test, it was shown that the forward voltage drop of the diode, at a constant current, has good linearity for neutron dosage. Also it was found that the newton dosage can be measured by the pin diode neutron dosimeter with constant current power.

Keywords