대한전자공학회:학술대회논문집 (Proceedings of the IEEK Conference)
- 대한전자공학회 2000년도 하계종합학술대회 논문집(1)
- /
- Pages.69-72
- /
- 2000
고속 디지털 MRI 모뎀 수신기 설계
Design of Receiver in High-Speed digital Modem for High Resolution MRI
초록
This paper shows the more improved design of MRI receiver compared to conventional one based on Elscint Spectrometer. At first, the low-cost ADC is 16 bits, 3MHz sampling A/D converter Comparing to conventional one with signal bits of 14 bits, this device with those of 16 bits helps getting Improved the image resolution improved. If frequency is designed centering around 7.6 MHz to be satisfied in 10 MHz of maximum input bandwidth of ADC. For 1st demodulation, fixed IF is used for the purpose of the implementing multi nuclei system. Control parts & partial digital parts are integrated on one chip(FPGA). In DDC(Digital Down Converter), we got required bandwidth of LPF by controlling its decimation rate. With above considerations, we designed optimal receiver for high resolution imaging to be implemented through PC interface & experimental test of receiver of MRI after receiver's fabrication.
키워드