Design and Evaluation of Uncertainty for 6-component Force/Moment Calibration Machine

6분력 힘/모멘트 교정기의 설계 및 불확도 평가

  • Published : 1997.04.01

Abstract

This paper presents the design and the evaluation of the 6-component force/moment calibration machine which h a s t h e maximum capacities of 500 N in forces and 50 Nm in moments. This calibration machine consists of body. fixture. force generating system, moment generating system. The expanded uncertainty of the calibration machine is evaluated by calculating the A type uncertainty. $U_A$ and B type uncertainty, $U_B$. The evaluation results. this system has the expanded uncertainty of less than $2{\times}10^[-2]$ in respective force and moment components.

Keywords