Development of Computerized Spirometer

Computerized spirometer의 개발

  • Cha, E.J. (Department of Biomedical Engineering, Chungbuk National University) ;
  • Park, I.S. (Department of Biomedical Engineering, Chungbuk National University) ;
  • Song, C.H. (Department of Biomedical Engineering, Chungbuk National University) ;
  • Kim, D.W. (Department of Biomedical Engineering, Chungbuk National University) ;
  • Goo, Y.S. (Department of Physiology, Chungbuk National University) ;
  • Lee, T.S. (Department of Biomedical Engineering, Chungbuk National University)
  • 차은종 (충북대학교 의과대학 의공학교실) ;
  • 박인석 (충북대학교 의과대학 의공학교실) ;
  • 송춘희 (충북대학교 의과대학 의공학교실) ;
  • 김도우 (충북대학교 의과대학 의공학교실) ;
  • 구용숙 (충북대학교 의과대학 생리학교실) ;
  • 이태수 (충북대학교 의과대학 의공학교실)
  • Published : 1996.11.15

Abstract

Computerized spirometer was developed in a form of proto type system. The system consisted of pneumotachometer, flow measurement unit, fan ventilator unit, and software program. Patient's respiratory flow signal was first converted to corresponding pressure drop signal by a screen type pneumotachometer, sensed by a differential pressure transducer, amplified and low pass filtered, and digitized at a rate of 100 Hz, then fed into a PC thru RS-232C serial port. Customized application software controls data acquisition followed by computation of test parameters. The fan ventilator unit dries and eliminates microorganism in the pneumotachometer after each test. The system performs conventional spirometic tests and manages the test results in a database for retrograde research. The proto type system was fully developed and the commercialized system is currently being built up.