Functional Assessment of Vestibular System and Dizziness Diagnosis

전정기능 평가 및 질병 진단을 위한 정현파 회전자극기 개발

  • Jeong, Ho-Chun (Dept. of Electronics Engineering, Wonkwang University) ;
  • Lim, Seung-Kwan (Dept. of Electronics Engineering, Wonkwang University) ;
  • Kim, Kuy-Kem (Dept. of Electronics Engineering, Wonkwang University) ;
  • Chin, Dal-Bok (Dept. of Electronics Engineering, Wonkwang University)
  • Published : 1996.07.22

Abstract

The purpose of this study was to develop systematic diagnostic system testing easily, rapidly vestibular function of patients suffered from vestibular syndrome such as nausea vomiting, dizzness, ataxia. Diagnostic system composed of rotatory chair system which rotated sinusoidally patients against their vertical axis for purpose of invoking eye movement by vestibulo-ocular reflex and the softwares which storaged eye movement into computer and analyzed eye movement. Rotatory chair system consisted of comfortable chair and DC servomotor with reducer(1:80) by controlled servo in field of nonlinear motor control, double feedback loops system containing velocity feedback loop and position feedback loop was applied to this sever controlled rotatory chair system. Maximum rotatory velocity of rotatory chair was upto 60 degree per second and frequency range was 0.01 to 0.64 Hz. These above results suggest that clinical rotatory chair system may test easily, rapidly vestibular function and diagnose etiology of dizziness, thus giving effective assistance on the treatment of dizziness patients.

Keywords