Proceedings of the KOSOMBE Conference (대한의용생체공학회:학술대회논문집)
- Volume 1991 Issue 05
- /
- Pages.5-7
- /
- 1991
A STATIC IMAGE RECONSTRUCTION ALGORITHM IN ELECTRICAL IMPEDANCE TOMOGRAPHY
임피던스 단층촬영기의 정적 영상 복원 알고리즘
- Woo, Eung-Je (Dept. of Biomed. Eng., Kon Kuk University) ;
- Webster, John G. (Dept. of Elec. and Comp. Eng., University of Wisconsin-Madison) ;
- Tompkins, Willis J. (Dept. of Elec. and Comp. Eng., University of Wisconsin-Madison)
- 우응제 (건국대학교 의과대학 의학공학과) ;
- ;
- Published : 1991.05.15
Abstract
We have developed an efficient and robust image reconstruction algorithm for static impedance imaging. This improved Newton-Raphson method produced more accurate images by reducing the undesirable effects of the ill-conditioned Hessian matrix. We found that our electrical impedance tomography (EIT) system could produce two-dimensional static images from a physical phantom with 7% spatial resolution at the center and 5% at the periphery. Static EIT image reconstruction requires a large amount of computation. In order to overcome the limitations on reducing the computation time by algorithmic approaches, we implemented the improved Newton-Raphson algorithm on a parallel computer system and showed that the parallel computation could reduce the computation time from hours to minutes.