• Title/Summary/Keyword: water phantom

Search Result 372, Processing Time 0.026 seconds

Evaluation of Dosimetric Characteristics of Small Field in Cone Versus Square Fields Based on Linear Accelerators(LINAC) for Stereotactic Radiosugery(SRS) (선형가속기를 기반으로 한 뇌정위 방사선 수술 시 전용 콘과 정방형 소조사면의 선량 특성에 관한 고찰)

  • Yoon, Joon;Lee, Gui-Won;Park, Byung-Moon
    • Journal of radiological science and technology
    • /
    • v.33 no.1
    • /
    • pp.61-66
    • /
    • 2010
  • In this paper we evaluated small field dose characteristics of exclusive cone fields versus square fields for stereotactic radiosugery (SRS) which is based on linear accelerators (LINAC). For this test, we used a small beam detector (stereotactic fields detector : SFD) with a 6 MV photon beam and a water phantom system (IBA, Germany). Percentage depth dose (PDD) was measured for different field sets (cones : ${\Phi}1\;cm$, ${\Phi}2\;cm$, ${\Phi}3\;cm$ ; square fields : $1{\times}1\;cm^2$, $2{\times}2\;cm^2$, $3{\times}3\;cm^2$) at a source skin distance (SSD) of 100 cm. We measured the point depths at 1.5 cm, 5 cm, 10 cm, 20 cm, and 30 cm. The output factors were measured under the same geometrical conditions of the PDD and normalized at the maximum dose depth. To analyze the penumbra, we measured the dose profile with 95 cm of SSD, 5 cm of depth for each field sizes (${\Phi}1\;cm$, ${\Phi}3\;cm$, $1{\times}1\;cm^2$, and $3{\times}3\;cm^2$) using SFD. We obtained the values for every 1 mm interval in the physical field (90%) and 0.5 mm interval in the penumbra region (20 to 80%). The PDD variation of exclusive cones and square fields were 4.3 to 7.9% lesser than the standard field size ($10{\times}10\;cm^2$. The variation of PDD was reduced while the field size was increased. To compare the beam quality, we analyzed the $PDD_{20,10}$ and the results showed under the 1% of variations for all experiments except for ${\Phi}1\;cm$ cone and $1{\times}1\;cm^2$ fields. Output factors of exclusive cone were increased 3.1~4.6% than the square fields, and the penumbra region of exclusive cone was reduced 20% as compared to the square fields. As the previous researches report, it is very important for SRS and SFD that precise dosimetry in small beam fields. In this paper, we showed the effectiveness of exclusive cone, compared to square field. And we will study on the various detector characteristics for small beam fields.

An Assessment of Post-Injection Transmission Measurement for Attenuation Correction With Rotating Pin Sources in Positron Emission Tomography (양전자방출단층촬영(PET)에서 회전 핀선원과 투과 및 방출 동시 영상 방법을 이용한 감쇠보정 방법 특성에 관한 고찰)

  • Lee, J.R.;Choi, Y.;Lee, K.H.;Kim, S.E.;Chi, D.Y.;Shin, S.A.;Kim, B.T.
    • The Korean Journal of Nuclear Medicine
    • /
    • v.29 no.4
    • /
    • pp.533-540
    • /
    • 1995
  • Attenuation correction is important in producing quantitative positron emission tomography (PET) images. Conventionally, photon attenuation effects are corrected using transmission measurements performed before tracer administration. The pre-injection transmission measurement approach may require a time delay between transmission and emission scans for the tracer studies requiring a long uptake period, about 45 minutes for F-18 deoxyglucose study. The time delay will limit patient throughput and increase the likelihood of patient motion. A technique lot performing simultaneous transmission and emission scans (T+E method) after the tracer injection has been validated. The T+E method substracts the emission counts contaminating the transmission measurements to produce accurate attenuation correction coefficients. This method has been evaluated in experiments using a cylindrical phantom filled with background water (5750 cc) containing $0.4{\mu}Ci/cc$ of F-18 fluoride ion and one insert cylinder (276 cc) containing $4.3{\mu}Ci/cc$. GE $Advance^{TM}$ PET scanner and Ge-68 rotating pin sources for transmission scanning were used for this investigation. Post-injection transmission scan and emission scan were peformed alternatively over time. The error in emission images corrected using post-infection transmission scan to emission images corrected transmission scan was 2.6% at the concentration of $1.0{\mu}Ci/cc$. No obvious differences in image quality and noise were apparent between the two images. The attenuation correction can be accomplished with post-injection transmission measurement using rotating pin sources and this method can significantly shorten the time between transmission and omission scans and thereby reduce the likelihood of patient motion and increase scanning throughput in PET.

  • PDF

Dose distribution at junctional area for head and neck radiotherapy (두경부 방사선치료시 접합 조사면의 선량분포)

  • 김정기;김기환;오영기;김진기;정동혁;신교철;양광모;조문준;박인규
    • Progress in Medical Physics
    • /
    • v.12 no.2
    • /
    • pp.161-169
    • /
    • 2001
  • For the head and neck radiotherapy, the technique of half beam using independent collimator is very useful to avoid overlapping of fields particularly when the lateral neck fields are placed adjacent to anterior supraclavicular field. Also abutting photon field with electron field is frequently used for the irradiation of posterior neck when tolerable dose on spinal cord has been reached. Using 6 MV X-ray and 9 MeV electron beams of Clinac1800(Varian, USA) linear accelerator, we performed film dosimetry by the X-OMAT V film of Kodak in solid water phantom and the dose distribution at beam center of 2 half beams further examined according to depths(0 cm, 1.5 cm, 3 cm, 5 cm) for single anterior half beam and anterior/posterior half beam. The dose distribution to the junction line between photon and electron fields was also measured. For the single anterior half beam, the absorption doses at 0.3 cm, 0.5 cm and 1 cm distances from beam center were 88%, 93% and 95% of open beam, respectively. In the anterior/posterior half beams, the absorption doses at 0.3 cm, 0.5 cm and 1 cm distances from beam center were 92%, 93% and 95% of open beam, respectively At the junction line between photon and electron fields, hot spot was developed on the side of the photon field and a cold spot was developed on that of the electron field. The hot spot in the photon side was developed at depth 1.5 cm with 7 mm width. The maximum dose of hot spot was increased to 6% of reference doses in the photon field. The cold spot in the electron side was developed at all measured depths(0.5 cm-3 cm) with 1-12.5 mm widths. The decreased dose in the cold spot was 4.5-30% of reference dose in the electron field. With above results, we concluded that when using electron beam or independent jaw for head and neck radiotherapy, the hot and cold dose area should be considered as critical point.

  • PDF

Commissionning of Dynamic Wedge Field Using Conventional Dosimetric Tools (선량 중첩 방식을 이용한 동적 배기 조사면의 특성 연구)

  • Yi Byong Yong;Nha Sang Kyun;Choi Eun Kyung;Kim Jong Hoon;Chang Hyesook;Kim Mi Hwa
    • Radiation Oncology Journal
    • /
    • v.15 no.1
    • /
    • pp.71-78
    • /
    • 1997
  • Purpose : To collect beam data for dynamic wedge fields using conventional measurement tools without the multi-detector system, such as the linear diode detectors or ionization chambers. Materials and Methods : The accelerator CL 2100 C/D has two photon energies of 6MV and 15MV with dynamic wedge an91es of 15o, 30o, 45o and 60o. Wedge transmission factors, percentage depth doses(PDD's) and dose Profiles were measured. The measurements for wedge transmission factors are performed for field sizes ranging from $4\times4cm^2\;to\;20\times20cm^2$ in 1-2cm steps. Various rectangular field sizes are also measured for each photon energy of 6MV and 15MV, with the combination of each dynamic wedge angle of 15o 30o. 45o and 60o. These factors are compared to the calculated wedge factors using STT(Segmented Treatment Table) value. PDD's are measured with the film and the chamber in water Phantom for fixed square field. Converting parameters for film data to chamber data could be obtained from this procedure. The PDD's for dynamic wedged fields could be obtained from film dosimetry by using the converting parameters without using ionization chamber. Dose profiles are obtained from interpolation and STT weighted superposition of data through selected asymmetric static field measurement using ionization chamber. Results : The measured values of wedge transmission factors show good agreement to the calculated values The wedge factors of rectangular fields for constant V-field were equal to those of square fields The differences between open fields' PDDs and those from dynamic fields are insignificant. Dose profiles from superposition method showed acceptable range of accuracy(maximum 2% error) when we compare to those from film dosimetry. Conclusion : The results from this superposition method showed that commissionning of dynamic wedge could be done with conventional dosimetric tools such as Point detector system and film dosimetry winthin maximum 2% error range of accuracy.

  • PDF

Effect of an Acrylic Plate and SSD on Dose Profile and Depth Dose Distribution of 9 MeV Electron Beams (에너지 저하체로서 아크릴과 SSD 가 9MeV 전자선의 측방 및 깊이선량분포에 미치는 효과)

  • 강위생
    • Progress in Medical Physics
    • /
    • v.9 no.2
    • /
    • pp.65-71
    • /
    • 1998
  • The aims are to evaluate the effects of an 1.0 cm acrylic plate and SSD on the dose profile and depth dose distribution of 9 MeV electron beam and to analyse adequacy for using an acrylic plate to reduce energy of electron beams. An acrylic plate of 1.0 cm thickness was used to reduce energy of 9 MeV electron beam to 7 MeV. The plate was put on an electron applicator at 65.4 cm distance from x-ray target. The size of the applicator was 10${\times}$l0cm at 100 cm SSD. For 100cm, l05cm and 110cm SSD, depth dose on beam axis and dose profiles at d$\_$max/ on two principal axes were measured using a 3D water phantom. From depth dose distributions, d$\_$max/, d$\_$85/, d$\_$50/ and R$\_$p/, surface dose, and mean energy and peak energy at surface were compared. From dose profiles flatness, penumbra width and actual field size were compared. For comparison, 9 MeV electron beams were measured. Surface dose of 7 MeV electron beams was changed from 85.5% to 82.2% increasing SSD from 100 cm to 110 cm, and except for dose buildup region, depth dose distributions were independent of SSD. Flatness of 7 MeV ranged from 4.7% to 10.4% increasing SSD, comparing 1.4% to 3.5% for 9 MeV. Penumbra width of 7 MeV ranged from 1.52 cm to 3.03 cm, comparing 1.14 cm to 1.63 cm for 9 MeV. Actual field size increased from 10.75 cm to 12.85 cm with SSD, comparing 10.32 cm to 11.46 cm for 9 MeV. Virtual SSD's of 7 and 9 MeV were respectively 49.8 cm and 88.5cm. In using energy reducer in electron therapy, depth dose distribution were independent of SSD except for buildup region as well as open field. In case of using energy reducer, increasing SSD made flatness to deteriorate more severely, penumbra width more wide, field size to increase more rapidly and virtual SSD more short comparing with original electron beam. In conclusion, it is desirable to use no energy reducer for electron beam, especially for long SSD.

  • PDF

Quantitative Analysis of Digital Radiography Pixel Values to absorbed Energy of Detector based on the X-Ray Energy Spectrum Model (X선 스펙트럼 모델을 이용한 DR 화소값과 디텍터 흡수에너지의 관계에 대한 정량적 분석)

  • Kim Do-Il;Kim Sung-Hyun;Ho Dong-Su;Choe Bo-young;Suh Tae-Suk;Lee Jae-Mun;Lee Hyoung-Koo
    • Progress in Medical Physics
    • /
    • v.15 no.4
    • /
    • pp.202-209
    • /
    • 2004
  • Flat panel based digital radiography (DR) systems have recently become useful and important in the field of diagnostic radiology. For DRs with amorphous silicon photosensors, CsI(TI) is normally used as the scintillator, which produces visible light corresponding to the absorbed radiation energy. The visible light photons are converted into electric signal in the amorphous silicon photodiodes which constitute a two dimensional array. In order to produce good quality images, detailed behaviors of DR detectors to radiation must be studied. The relationship between air exposure and the DR outputs has been investigated in many studies. But this relationship was investigated under the condition of the fixed tube voltage. In this study, we investigated the relationship between the DR outputs and X-ray in terms of the absorbed energy in the detector rather than the air exposure using SPEC-l8, an X-ray energy spectrum model. Measured exposure was compared with calculated exposure for obtaining the inherent filtration that is a important input variable of SPEC-l8. The absorbed energy in the detector was calculated using algorithm of calculating the absorbed energy in the material and pixel values of real images under various conditions was obtained. The characteristic curve was obtained using the relationship of two parameter and the results were verified using phantoms made of water and aluminum. The pixel values of the phantom image were estimated and compared with the characteristic curve under various conditions. It was found that the relationship between the DR outputs and the absorbed energy in the detector was almost linear. In a experiment using the phantoms, the estimated pixel values agreed with the characteristic curve, although the effect of scattered photons introduced some errors. However, effect of a scattered X-ray must be studied because it was not included in the calculation algorithm. The result of this study can provide useful information about a pre-processing of digital radiography.

  • PDF

Synthesis of Ultrasound Contrast Agent: Characteristics and Size Distribution Analysis (초음파 조영제의 합성 및 합성된 초음파 조영제의 특성 분석)

  • Lee, Hak Jong;Yoon, Tae Jong;Yoon, Young Il
    • Ultrasonography
    • /
    • v.32 no.1
    • /
    • pp.59-65
    • /
    • 2013
  • Purpose: The purpose of this study is to establish the methodology regarding synthesis of ultrasound contrast agent imaging, and to evaluate the characteristics of the synthesized ultrasound contrast agents, including size or degradation interval and image quality. Materials and Methods: The ultrasound contrast agent, composed of liposome and SF6, was synthesized from the mixture solution of $21{\mu}mol$ DPPC (1, 2-Dihexadecanoyl-sn-glycero-3-phosphocholine, $C_{40}H_{80}NO_8P$), $9{\mu}mol$ cholesterol, $1.9{\mu}mol$ of DCP (Dihexadecylphosphate, $[CH_3(CH_2)_{15}O]_2P(O)OH$), and chloroform. After evaporation in a warm water bath and drying during a period of 12-24 hours, the contrast agent was synthesized by the sonication process by addition of buffer and SF6 gas. The size of the contrast agent was controlled by use of either extruder or sonication methods. After synthesis of contrast agents, analysis of the size distribution of the bubbles was performed using dynamic light scattering measurement methods. The degradation curve was also evaluated by changes in the number of contrast agents via light microscopy immediate, 12 hours, 24 hours, 36 hours, 48 hours, 60 hours, 72 hours, and 84 hours after synthesis. For evaluation of the role as an US contrast agent, the echogenicity of the synthesized microbubble was compared with commercially available microbubbles (SonoVue, Bracco, Milan, Italy) using a clinical ultrasound machine and phantom. Results: The contrast agents were synthesized successfully using an evaporation-drying-sonication method. The majority of bubbles showed a mean size of 154.2 nanometers, and they showed marked degradation 24 hours after synthesis. ANOVA test revealed a significant difference among SonoVue, synthesized contrast agent, and saline (p < 0.001). Although no significant difference was observed between SonoVue and the synthesized contrast agent, difference in echogenicity was observed between synthesized contrast agent and saline (p < 0.01). Conclusion: We could synthesize ultrasound contrast agents using an evaporation-drying-sonication method. On the basis of these results, many prospective types of research, such as anticancer drug delivery, gene delivery, including siRNA or microRNA, targeted molecular imaging, and targeted therapy can be performed.

Effects of Gadolinium Contrast agent on Bone Mineral Density Measurement using Dual Energy X-ray Absorptiometry (가돌리늄조영제가 이중에너지 X-선 흡수법을 이용한 골밀도검사에 미치는 영향)

  • Lee, Keun-Ohk;Lee, Min-Su
    • Journal of the Korean Society of Radiology
    • /
    • v.15 no.1
    • /
    • pp.63-70
    • /
    • 2021
  • Radiographic contrast agents are used for diagnostic purposes and are one of the factors affecting measured values in bone density tests. They are absorbed into tissues and have an effect of increasing the measured values of bone density, so they are avoided as much as possible before performing a bone density test. MRI contrast agents, which have different physical properties and mechanisms of action than radiographic contrast agents, are based on gadolinium, a metal element. They have radiopacity characteristics, so MRI are generally performed prior to examination using radiation. The purpose of this study was to investigate the effects of MRI contrast agents on bone mineral density examination using dual energy X-ray absorption. Two types of gadolinium based MRI contrast agents were injected into an acrylic water tank for each volume, and the humanoid spine phantom was inserted and the BMD and T-score from (L1-L4) were analyzed by scanning a total of 30 times, 5 times for each injection type. The average value of the measured total (L1-L4) bone density for each of the two contrast agents was 0.952±0.052, 0.957±0.050, and 0.956±0.05g/㎠, respectively, for the Gadoterate Meglumine component 0mL, 7.5mL and 15mL, when the gadobutrol components were 0mL, 5mL, and 10mL, there was no statistically significant difference at all sites at 0.953±0.001, 0.954±0.001, and 0.945±0.001g/㎠, respectively(p>0.05). The average value of total T-score was -0.46±0.05, -0.4±0, -0.42±0.04 when the Gadoterate Meglumine component was 0mL, 7.5mL and 15mL, respectively. When the Gadobutrol ingredients were 0mL, 5mL and 10mL, there was no statistically significant difference in all areas, with -0.46±0.05, -0.46±0.05, and 0.5±0.00, respectively. In this experiment, the MRI contrast agent was found to have no effect on bone density tests, using the dual-energy X-ray absorption method. There is a limitation in that physical conditions such as kidney and health conditions etc. were not taken into consideration, so further clinical research is expected to be conducted in the future.

Dosimetry of the Low Fluence Fast Neutron Beams for Boron Neutron Capture Therapy (붕소-중성자 포획치료를 위한 미세 속중성자 선량 특성 연구)

  • Lee, Dong-Han;Ji, Young-Hoon;Lee, Dong-Hoon;Park, Hyun-Joo;Lee, Suk;Lee, Kyung-Hoo;Suh, So-Heigh;Kim, Mi-Sook;Cho, Chul-Koo;Yoo, Seong-Yul;Yu, Hyung-Jun;Gwak, Ho-Shin;Rhee, Chang-Hun
    • Radiation Oncology Journal
    • /
    • v.19 no.1
    • /
    • pp.66-73
    • /
    • 2001
  • Purpose : For the research of Boron Neutron Capture Therapy (BNCT), fast neutrons generated from the MC-50 cyclotron with maximum energy of 34.4 MeV in Korea Cancer Center Hospital were moderated by 70 cm paraffin and then the dose characteristics were investigated. Using these results, we hope to establish the protocol about dose measurement of epi-thermal neutron, to make a basis of dose characteristic of epi-thermal neutron emitted from nuclear reactor, and to find feasibility about accelerator-based BNCT. Method and Materials : For measuring the absorbed dose and dose distribution of fast neutron beams, we used Unidos 10005 (PTW, Germany) electrometer and IC-17 (Far West, USA), IC-18, ElC-1 ion chambers manufactured by A-150 plastic and used IC-l7M ion chamber manufactured by magnesium for gamma dose. There chambers were flushed with tissue equivalent gas and argon gas and then the flow rate was S co per minute. Using Monte Carlo N-Particle (MCNP) code, transport program in mixed field with neutron, photon, electron, two dimensional dose and energy fluence distribution was calculated and there results were compared with measured results. Results : The absorbed dose of fast neutron beams was $6.47\times10^{-3}$ cGy per 1 MU at the 4 cm depth of the water phantom, which is assumed to be effective depth for BNCT. The magnitude of gamma contamination intermingled with fast neutron beams was $65.2{\pm}0.9\%$ at the same depth. In the dose distribution according to the depth of water, the neutron dose decreased linearly and the gamma dose decreased exponentially as the depth was deepened. The factor expressed energy level, $D_{20}/D_{10}$, of the total dose was 0.718. Conclusion : Through the direct measurement using the two ion chambers, which is made different wall materials, and computer calculation of isodose distribution using MCNP simulation method, we have found the dose characteristics of low fluence fast neutron beams. If the power supply and the target material, which generate high voltage and current, will be developed and gamma contamination was reduced by lead or bismuth, we think, it may be possible to accelerator-based BNCT.

  • PDF

The Study of Shielding Effect on Ovoids of Three Different Gynecological Applicator Sets in microSelectron-HDR System (microSelectron-HDR System에서 부인암 강내조사에 쓰이는 세 가지 Applicator Set들의 Ovoids에 대한 차폐효과 연구)

  • Cho, Young-K.;Park, Sung-Y.;Choi, Jin-H.;Kim, Hung-J.;Kim, Woo-C.;Loh, John-J.K.;Kim, Joo-Y.
    • Journal of Radiation Protection and Research
    • /
    • v.23 no.4
    • /
    • pp.259-266
    • /
    • 1998
  • There are three different types of gynecological applicator sets available in microSelectron-high dose-rate(HDR) System by Nucletron; standard applicator set(SAS), standard shielded applicator set(SSAS), and Fletcher-Williamson applicator set(FWAS). Shielding effect of a SAS without shielding material was compared with that of a SSAS with shielding material made of stainless steel(density ${\varrho}=8,000kg/m^3$) at the top and bottom of each ovoid, and of a FWAS with shielding material made of tungsten alloy(density ${\varrho}=14,000kg/m^3$ at the top and bottom of each ovoid. The shielding effects to the rectum and bladder of these two shielded applicator sets were to be measured at reference points with an ion chamber and specially designed supporting system for applicator ovoids inside of the computerized 3-dimensional water phantom. To determine the middle point of two ovoids the measurement was performed with the reference tip of ion chamber placed at the same level and at the middle point from the two ovoids, while scanning the dose with the ion chamber on each side of ovoids. The doses to the reference points of rectum were measured at 20(Rl), 25(R2), 30(R3), 40(R4), 50(R5), and 60(R6) mm located posteriorly on the vertical line drawn from M5(the middle dwell position of ovoid), and the doses to the bladder were measured at 20(Bl), 30(B2), 40(B3), 50(B4), and 60(B5) mm located anteriorly on the vertical line drawn from M5. The same technique was employed to measure the doses on each reference point of both SSAS and FWAS. The differences of measured rectal doses at 25 mm(R2) and 30 mm(R3) between SAS and SSAS were 8.0 % and 6.0 %: 25.0% and 23.0 % between SAS and FWAS. The differences of measured bladder doses at 20 mm(Bl) and 30 mm(B2) between SAS and SSAS were 8.0 % and 3.0 %: 23.0 % and 17.0 % between SAS and FWAS. The maximum shielding effects to the rectum and bladder of SSAS were 8.0 % and 8.0 %, whereas those of FWAS were 26.0 % and 23.0 %, respectively. These results led to the conclusion that FWAS has much better shielding effect than SSAS does, and when SSAS and FWAS were used for gynecological intracavitary brachytherapy in microSelectron-HDR system, the dose to the rectum and bladder was significantly reduced to optimize the treatment outcome and to lower the complication rates in the rectum and bladder.

  • PDF